3.4 Relative Display Function
Canceling the Relative
function:
Press
H.AUTO/REL
1.
Move the function selector
to the
Ω
position and con-
nect the test leads to the
equipment.
2.
Short-circuit the test leads.
The resistance of the test
leads is displayed.
3.
Press the
H.AUTO/REL
button for at least 1 sec-
ond.
("REL" lights up, "0000")
4.
Connect the test leads to
the measurement object,
and read the indicated
value.
Application: Using the Relative Display function for the purpose
of zero adjustment in the measurement of low resistances
H.AUTO/REL
Measure the resistance.
for at least
1 second
Short-circuit the test leads