![Hioki 2331-20 Instruction Manual Download Page 52](http://html1.mh-extra.com/html/hioki/2331-20/2331-20_instruction-manual_2137160052.webp)
5.3 General Specifications
48
5.3 General Specifications
Input Method
U INPUT: Insulated input
(Not insulated from current measurement
circuit)
CLAMP SENSOR INPUT:
Input insulated by a clamp sensor
Input Resistance
(50/60 Hz)
U INPUT: 1.6 M
Ω
±
10% (Difference input)
CLAMP SENSOR INPUT: 200 k
Ω
±
10%
Measurement
Method
Digital sampling
Internal Memory 512 k bytes Flash memory
Backup
Recorded data (saved in flash memory)
Data loss for up to 2 minutes before and
after a power outage may occur.
Clock Function
The real time clock (year, month, day, hour,
minute, and second) of the communications
module is used.
LED Display
Used for monitoring and warning
Communication
Interface
CAN bus
Maximum Input
Voltage
U INPUT: 300 Vrms, 424.3 V peak
ALARM : 30 VDC
CLAMP SENSOR INPUT: 1.5 Vrms, 2.2 V
peak value
Maximum Rated
Voltage to Earth
Voltage input (U INPUT) terminal
300 Vrms, 50/60 Hz
Rated Supply
Voltage
+5V±0.3 VDC
Maximum Rated
Power
2.5 W
Summary of Contents for 2331-20
Page 1: ...INSTRUCTION MANUAL POWER METER 2331 20 MODULE ...
Page 2: ......
Page 26: ...1 4 Dimension Diagrams 22 ...
Page 28: ...2 1 Setting the Module ID 24 ...
Page 44: ...4 2 Insulation of Internal Circuit 40 ...
Page 54: ...5 3 General Specifications 50 ...
Page 56: ...6 2 Service 52 ...
Page 57: ......
Page 58: ......