29
Electromagnetic Compatibility
Medical electrical equipment needs special precautions regarding electromagnetic
compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in this user manual.
Portable and mobile radio frequency (RF) communications equipment can affect medical
electrical equipment.
Electromagnetic Emissions Guidance
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The Basic Care™ Bed is intended for use in the electromagnetic environment specified
below. The customer or the user of the bed should make sure it is used in such an
environment.
Emissions Test
Compliance
Electromagnetic Environment—Guidance
RF Emissions
CISPR 11
Group 1
The Basic Care™ Bed uses RF energy only for its
internal functions. Therefore, its RF emissions are
low and are not likely to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The Basic Care™ Bed is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Not
applicable
Voltage
Fluctuations/
Flicker Emissions
IEC 61000-3-3
Not
applicable