1 4 Technical data
55000-00012_002_01 – 1695233 – 25/01/2018
53
14.2
EMC information
EMC notes
Operate the device only with the stated accessories.
TruLight™1000 may only be operated with the accessories stated in the accompanying documents. Operation with
accessories, converters or conduits other than those stated in the accompanying documents can lead to increased
EMC emissions or reduced interference immunity of the device and thus to improper use.
Install or operate the device according to the EMC instructions
Medical electrical devices such as the TruLight™ 1000 are subject to specific precautionary measures regarding EMC
and must be installed and operated in accordance with the EMC instructions stated in the accompanying
documents.
Do not stack the device with or next to other devices
TruLight™ 1000 must not be stacked immediately next to or on top of other devices.
WARNING
NOTE
NOTE
Guidelines and manufacturer's declaration -
electromagnetic immunity
TruLight™ 1000 is intended for use in the ELECTROMAGNETIC ENVIRONMENT described below. The customer or user should
ensure that the TruLight™ 1000 is operated in an environment as described.
Emission measurements
Compliance
ELECTROMAGNETIC ENVIRONMENT - guidelines
RF emissions in accordance with CISPR 11 Group 1
TruLight™ 1000 uses RF energy exclusively for its internal
FUNCTION. Therefore, its RF emission levels are very low,
and it is unlikely that neighbouring electronic devices
would be adversely affected.
RF emissions in accordance with CISPR 11 Class B
TruLight™ 1000 is suitable for use in all establishments,
including residential areas and facilities that are directly
connected to a PUBLIC POWER SUPPLY NETWORK which
also supplies buildings that are used for residential
purposes.
Harmonic emissions in accordance with
EN / IEC 61000-3-2
Class A
Voltage fluctuations / flicker emissions
in accordance with EN / IEC 61000-3-3
Fulfilled