Commissioning
Figure 9
6.1
Checking the lighting system
For all versions of the lighting system, a function test and a
visual inspection should be carried out before use, or at least
once per week.
Contamination and infection hazard for
patients
Loose or damaged parts may fall into wounds. To
ensure the safety of patients, check the
components of the control screen for the following
points before each use:
• Loose parts on light head ①,
• visible damage, in particular on the cover
plates ② of the light head and the sterilisable
handle ③,
• secure moun ng of the sterilisable handle ③.
------------------------------------------------------------
Electric shock
There is a risk of electric shock in the
event of contact with damaged
electrical components of the wall-mounted /
mobile stand version:
• Do not connect the lighting system to the
mains in the event of defective plug
connectors ④ or ⑥ or damaged mains
cables ⑤.
The lighting system is no longer safe to use when the
damage described above or other damage occurs:
• Disconnect the lighting system by using the master switch
installed in the building, or remove the safety connector.
• Secure the master switch or the safety connector against
accidental switch-on / plugging in.
• Label the lighting system as DEFECTIVE!
• Inform Technical Customer Service.
Strong magnetic fields
The support arm systems of the lighting systems
must not be used in the vicinity of strong magnetic
fields.
------------------------------------------------------------
BF / CF Class application components
No BF or CF Class application components in
accordance with IEC 60601-1 may be directly
WARNING