2 Safety instructions
55000-00012_002_01 – 1695233 – 25/01/2018
Swivel movement of the light head
Risk of injury due to uncontrolled swivel movement
Uncontrolled movement of the spring arm may result when
the spring force of the spring arm is not correctly adjusted.
---------------------------------------------------------------------------------------
Risk of jamming
When swivelling the light head, the distance between the
quarter bracket and the light head changes:
• Do not insert your fingers between the quarter bracket and the light
head when rotating the light head.
• Only grasp the light head with the sterilisable handle.
Cleaning and disinfection
Improperly used cleaning agents or disinfectants (see
Chapter 8
)
can pose a risk for patients or damage products
If the following information and instructions are not observed or
complied with, this may result in a risk of contamination or infection for
the patient or damage to the product. Furthermore, it would render any
claim for damages void!
• Use the wipe-over method only for disinfection.
• To clean or disinfect the device, the cloth for wiping must be moist
and not wet.
• Dispense cleaning agents and disinfectants so that no liquid can
enter through joints or openings of the surgical light or parts of the
support arm system.
• Use the surface disinfectant only at the concentration specified by
the manufacturer.
• Only use disinfectants approved by the manufacturer for use with
the following materials:
Polycarbonate (PC), polyamide (PA), acrylonitrile butadiene styrene
copolymer (ABS), polystyrene (PS), polyurethane (PUR),
polyphenylene sulphone (PPSU), polybutylene terephthalate (PBT)
and silicones.
• In the event of an increased layer formation of surface disinfectant,
thorough cleaning must be performed.
• Due to the risk of surface damage:
–
Do not use sharp, pointed or abrasive objects
–
Do not use abrasive substances or agents which can remove material
–
Do not use solvents, benzene, paint thinners, alkaline cleaning agents or
cleaning agents containing acids or aldehydes
–
Do not use agents with glycol derivatives, phenols, phenol derivatives or
quaternary compounds
WARNING
WARNING