Features
30
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
User Manual (143434 REV 8)
1. Scroll to Change delay. Press ENTER.
2. On the left side of the screen, scroll to either a 0, 2, 4, or 6-second alarm delay. Press ENTER. The filled
circle indicates the selected delay duration.
Bed Exit
1. Scroll to Bed exit: On/Off. Press ENTER.
2. Select either On to activate the Bed Exit detection feature or Off to cancel the Bed Exit functions.
Sound On/Off
1. Scroll to Sound On/Off. Press ENTER.
2. Select either On for an active audible indication or Off to cancel the audible indication. This only affects the
audible alarm on the bed. A nurse call is still placed.
NOTE:
If the bed does not have nurse call capabilities, the audible alarm is always active.
WARNING:
The Bed Exit System is not intended as a substitute for good nursing practices. The Bed Exit
System must be used in conjunction with a sound risk assessment and protocol.
WARNING:
The activation of 2, 4, or 6-second delay of the Bed Exit alarm feature may reduce the
effectiveness of the Bed Exit System.
NOTE:
If the bed is equipped with the pressure redistribution surface, the bed scale system will automatically re-calculate the
zero weight when adding or removing the pulmonary modules. There is no need to re-zero the scale.
Bed Setup/Reset
The Bed setup/Reset control clears the weight history and re-zeroes the scale.
Bed RESET
1. From the Home Screen, scroll to GO TO Next Screen. Press
ENTER.
2. Scroll to Bed setup/Reset. Press ENTER.
3. Scroll to Bed RESET. Press ENTER.
NOTE:
If the bed is equipped with a pressure redistribution surface, the Therapy Statistics will be cleared with Bed RESET.
Set Time and Date
1. Scroll to Set Time and Date. Press ENTER.
2. Move the up and down arrows to change the time and date. Press ENTER.