Technical Specifications
TotalCare® Bariatric Bed and TotalCare® Bariatric Plus Therapy System
99
User Manual (143434 REV 8)
Wireless Interface Unit
The WIU supports the following security protocols:
•
WEP (64 and 128 bit)
•
WPA - PSK with TKIP
•
WPA2 - PSK w/AES
•
WPA - 802.1x PEAP MSCHAPv2 w/TKIP
•
WPA2 - 802.1x PEAP MSCHAPv2 w/AES
b. Over the frequency range of 150 kHz to 80 MHz, field strength should be less than
3
V/m.
Recommended separation distances between portable and mobile RF
communications equipment and the P1840 Model
The P1840 is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the P1840 can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the P1840 as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter,
W
Separation distance according to frequency of transmitter, m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance
d
in metres (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.