7. The unit goes into Hyper-Inflate mode to inflate the mattress to 55 mm Hg (29"H
2
O)
for 13 minutes:
• If the unit is in Low Airloss mode,
the Hyper-Inflate screen shows.
•
If the unit is in Side Deflate mode,
the Side Deflate Active (Hyper-
Inflate Active) screen shows.
Deflate and Inflate the Side Bolsters
1. At the Low Airloss screen, press
Options. The Select Options screen
shows.
2. Press Deflate Sides. The Sides
Deflating screen shows. The side
bolsters deflate for four minutes, and
then the Side Deflate Active screen
shows.
NOTE:
During Side Deflate mode, you may use
Seat Deflate mode, Max Firm mode, or
Fowler Boost mode as necessary.
3. If the side bolsters are deflated for 30
minutes, an alarm sounds, and the
Warning: Side Deflate Active screen
shows. To turn the alarm off for 10
minutes, press Alarm Off.
4. To inflate the side bolsters, do as follows:
a. Press Options. The Select Options
screen shows.
b. Press Inflate Sides. The side
bolsters inflate, and the unit returns
to Low Airloss mode in 10 seconds.
28