Error message
A
Excess quality flushing duration has been exceeded
@
Even after quality flush, the permeate conductivity is still bigger than the set limit
value
A
LV Permeate Max
@
.
The error message is displayed and emergency service starts.
The yellow warning light
A
CDmax
@
additionally lights up.
Confirm the fault display with the
A
Alarm
@
key.
Check the settings of the plant and the raw water quality or check the units for possible defects.
The error display is only deleted after the fault has been eliminated.
Error message
A
Temperature low
@
The water temperature is lower than 0.5
C.
The water temperature during production has fallen below the minimum water temperature.
Check whether short circuiting has occurred in the sensor or the temperature sensor PT100 or in their supply lines.
Confirm the fault display with the
A
Alarm
@
key.
The error display is only deleted after the fault has been eliminated.
Error message
A
Emergency program / By-pass
@
The level in the storage tank is below Empty-level.
The signal occurs at contact input PROG and the input is programmed as
A
Emergency program BP
@
contact.
Confirm the fault display with the
A
Alarm
@
key.
Check the plant and, if necessary, its design as well as the
A
EMPTY
A
contact of the level controller. It is possible that it is
not sufficiently supplied if the production from the storage tank is too high.
The error display is only deleted after the fault has been eliminated.
Exc.Q-flush. time
196.2μS/cm 18°C
Temperature low
16.8μS/cm 0°C
Emerg.program /BP
16.8μS/cm 18°C