22.3 Autoclaving
Swing-out rotors, angle rotors made of aluminium and accessory parts can be autoclaved at 121°C (20 min).
The lids of the rotors and containers must be removed prior to autoclaving.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
22.4 Centrifuge
containers
•
With leakiness or after the breakage of centrifuging containers broken container parts and leaked centrifugation
material are to be completely removed.
•
The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splitters cause further glass breakage!
•
If this concerns infectious material, a disinfection process is to be executed immediately.
23 Faults
If the fault cannot be rectified according to the faults table, Hettich customer services must be informed.
Please state the type of centrifuge and the factory serial number. Both values are visible on the centrifuge type plate.
MAINS RESET:
- Mains switch OFF for longer than 10 secs.
- Mains switch ON.
Message / fault
Cause
Remedy
No display
---
−
No voltage.
−
Overvoltage protection tripped out.
−
Check supply voltage.
−
Mains switch ON.
01
−
Faulty speedometer.
TACHO - ERROR
02
−
No rotor installed.
−
Defective motor, frequency converter or
drive.
−
Open lid.
−
Turn rotor manually.
−
MAINS-RESET when power is
switched on, rotor should turn.
IMBALANCE ---
Imbalance
about
motor axis through weight
differential in rotor assembly.
−
Open lid.
−
Correct imbalance.
CONTROL - ERROR
04,
06 - 09
Error in lid locking or lid closure.
N > MAX
05
Rotation too fast
N < MIN
13
Rotation too slow
ROTORCODE
10
Incorrect rotor coding
−
Open lid.
−
MAINS-RESET
MAINS INTERRUPT
---
Power failure, centrifuging not properly
completed
−
Open lid.
−
Push
START
button.
VERSIONS-ERROR
12
Mismatch between electronic components
CONTROL-ERROR
21 - 27 Error / defective control unit
SER I/O - ERROR
30 - 38 Error / defective interface
FU / CCI - ERROR
60 - 83 Error / defective motor control
−
Open lid.
−
MAINS-RESET
N
>
ROTOR-MAX
---
Speed in the selected program greater than
the maximum speed of the rotor.
−
Check the set speed.
Reduce the set speed
24 Change mains input fuse
Switch off the mains switch and separate the centrifuge from the mains!
A
B
The fuse holder (A) with the mains input fuses is located next to the mains switch.
•
Remove the connecting cable from the machine plug socket.
•
Press the snap-fit (B) against the fuse holder (A) and remove.
•
Exchange defective mains input fuse.
•
Reinsert the fuse holder until the snap-fit clicks shut.
•
Reconnect the centrifuge to the mains supply.