19/34
5. Put on carefully the upper centrifuge housing.
6. Connect the centrifuge to the mains and switch the centrifuge on.
7. Switch on the mains switch (position "I").
The set machine version will be
displayed, e.g.:
RPM x 100
t
START
STOP
8. Adjust the machine version 3 by using the and keys beneath the time
indicator.
9. Press the key
STOP
in order to store the set machine version.
10. Switch off the mains switch (position "0").
11. Disconnect the centrifuge from mains and remove the short circuit plug.
12. Put on the upper part of the centrifuge housing and fix it.
13. Switch on the mains switch (position "I").
6.2
Enquiry the machine version
1. Switch off the mains switch.
2. Keep the key beneath the speed indicator and the key
pressed simultaneously.
3. Switch on the mains switch and release the keys again.
The speed indicator shows the machine version and the time indicator shows the set
brake step: e.g.:
RPM x 100
t
START
STOP
RPM x 100
t
START
STOP
If the machine version and brake step are not displayed, press the key under the
speed indicator until they are displayed.
4. To exit the machine version display press the key
STOP
or perform a MAINS-
RESET.