4 Troubleshooting
procedures
•
Fuses in installation in which centrifuge is installed are intact.
•
Mains input fuses of centrifuge are intact.
•
Supply voltage present at (see circuit diagram):
−
Connecting cable
−
Appliance plug
−
Mains switch
−
Power board A1, plug S402L and S402N
•
Look for the displayed error code in the chapter "Error messages".
•
Remedy the error according to the instructions.
•
Carry out a functional check after every repair and whenever a component is
replaced, see chapter "Functional check after a repair".