10
21/03/2019 Rev.2 (04/12/2019)
The lifting pole enables the patient to get out of the bed easily.
Beware: The maximum load capacity of the pole is 75kg at the front end.
Mounting the pole
In both corners of the backrest are two pole holders (1) with a notch which restricts the swivel range (2).
The pole has a metal pin which restricts the swivel range (3+4).
Insert the pole in the pole holder and check that the metal pin is in the notch of the pole holder.
Mounting the trapeze
The trapeze pole should be installed on the side where the patient gets in and out of bed. The pole can be moved on the left or right
sides.
Don‘t swivel the pole outside of the bed.
The trapeze (5) needs to be attached to the pole for use.
Pull the loop over the top of the pole (5) and adjust the height of the trapeze using the strap (from 55 cm to 70 cm) (6).
Make sure that the strap is positionned correctly through the buckle and tightly pull the trapeze downward to check it is secured.The
trapeze has a date clock (7). With normal use the trapeze will last for five years. After 5 years, replace the trapeze.
Step 6 :
Mounting and using the lifting pole.
1
2
3
Distributør i Norge:
Gymo AS - www.gymo.no
(+47) 4000 3522 - [email protected]