background image

 

   

Precision Balance                                                                                                                                                                                           

- 7 - 

 
 

Parameter Setting 

 

Hold down

key, then press

ON/OFF

key to enter the setting mode. The screen will show 

  : 

  key: 

To switch functions;

  key: 

To enter or exit

 

function setting selection;

key: 

To end setting.   

LCD Display 

  Function 

Selection   

Selection   

Page 

key: 

 

  key: 

key: 

 

 

Internal value 

display 

 

 

Item 1,   

Page     

 

Single point   

auto calibration 

 

 

Page 4 

 

Weight unit 

selection 

g, ct, lb, oz, ozt             
g, ct, lb, GN, dwt           
g, ct, lb, ozt, tl.j             
g, ct, lb, oz, tl.T             
g, ct, lb, g/m

2

, oz/yd

2   

 

g, ct, lb, XX, XX             

For application in Europe, America and China 
For application in Britain 
For application in Hong Kong 
For application in Taiwan 
For plaiting purpose 
For other purposes 

ON/OFF 

Item 2,   

Page 8 

 

Initial weight 

  unit selection 

g,  ct,  lb,  oz,  dr,  GN,  ozt,  dwt,  MM,  tl.j,  tl.T,  tl.H,  t,  oz/yd

2

,  g/m

2

 

Setting according to the specific applications in various regions 

 

Item 3,   

Page 8 

 

transfers rate 

selection 

9600 

 

Item 4, 

  Page 8 

 

Auto zero 

range selection 

d0/1/2/3/4/5 

 

Item 5,   

Page 9 

 

filter selection 

Fil1/2/4/8 

 

Item 6, 

  Page 9 

 

Backlight 

selection 

onoFF/oFF/ON 

 

Item 7, 

  Page 9 

 

Printing mode 

selection 

Pr-St/Co 

 

Item 8,   

Page 9 

 

External device 

PC

computer

/ SH-24(matrix printer)/

 

large screen 

 

Item 9, 

  Page 9 

Summary of Contents for HXB series

Page 1: ...PRECISION BALANCE USER S MANUAL MODEL X 2007 V2 0 ...

Page 2: ...eys 2 Functional Operation of Keys 3 Panel Illustration 3 Single Point Calibration of Weight 4 Single Point Calibration of Weight when Power on 5 Linearity Calibration of Weight 6 Parameter Setting 7 Parameter Functional Description 8 Units Conversion Table 10 Cross reference Table of Full Loads 10 RS 232 Serial Interface 11 ...

Page 3: ...en it clockwise Precautions 1 Please pay attention to the environment when performing weight calibration on the scale Avoid wind blow and vibration 2 Use of wrong batteries or wrong wiring may cause danger Descriptions of Keys ON OFF Power switch on the left rear side of the scale when it s in the upright position The zero key for resetting the equipment functional only when weight is within 4 of ...

Page 4: ...ounting Percentage Press key the screen will show then press key and the screen will show You may select sample size of 20 50 100 200 or 500 in turn by pressing in succession After selection put the sample on the weighing pan and start using the percentage function when you hear a beep sound Notes 1 Min unit weight of sample 1 d e g unit weight of sample of 300g is at least bigger than 0 01g 2 Whe...

Page 5: ... Step III Press key to start zero point calibration Make sure there is nothing on the weighing pan Step IV Please place weights of equivalent weight onto the pan Place 1 3 of load Place 2 3 of load Place full load Wait for a beep sound Step V Press key then press key Step VI After countdown the single point calibration is completed and the scale returns to the normal weighing mode ...

Page 6: ...Display Step I Hold down key Step II Then press key to start zero point calibration Make sure there is nothing on the weighing pan Step III Please place weights of equivalent weight onto the pan Place 1 3 of load Place 2 3 of load Place full load Wait for a beep sound Step IV Press key ...

Page 7: ... on the weighing pan and press to perform 1 3 load calibration Wait for a beep sound Step V Place 2 3 of load on the weighing pan and press to perform 2 3 load calibration Wait for a beep sound Step VI Place full load on the weighing pan and press to perform full load calibration Wait for a beep sound Step VII Clear the weighing pan then press key Step VIII After countdown the linearity calibratio...

Page 8: ...or application in Europe America and China For application in Britain For application in Hong Kong For application in Taiwan For plaiting purpose For other purposes ON OFF Item 2 Page 8 Initial weight unit selection g ct lb oz dr GN ozt dwt MM tl j tl T tl H t oz yd 2 g m 2 Setting according to the specific applications in various regions Item 3 Page 8 transfers rate selection 9600 Item 4 Page 8 A...

Page 9: ...displayed Press to switch between ON and OFF to use the unit displayed select on if you do not require the unit displayed select oFF Press to select from 5 kinds of weighing units after selecting the unit required press to return to the display 3 When the screen shows press to enter initial unit selection mode LCD display means the initial unit is g press to switch the unit required then Press key...

Page 10: ...s key to return to the display FiL1 2 4 8 The higher the value is the higher the filter level is 7 When the screen shows press to enter backlight setting mode The screen shows onoFF represents the auto backlight setting Press to select the desired backlight setting and press key to return to the display onoFF means auto backlight oFF means backlight off ON means backlight on 8 When the screen show...

Page 11: ...t matrix mini printer 10 When the screen shows press to enter display range selection mode LCD display represents display range Press to select the desired external device then press key to return to the display Zero0 1 2 3 4 5 means weight displayed when the weight exceeds the division 11 Before entering function setting press to end function setting and return to normal weighing mode Press the k...

Page 12: ...005 1 323X0 0001 3 3080X0 0002 6 6160X0 0005 Ounce Avoir 5 2926X0 0002 10 5855X0 0005 21 171X0 001 52 926X0 002 105 855X0 005 Dram Avoir 84 685X0 005 169 37X0 01 338 72X0 02 846 85X0 05 1693 7X0 1 Grain 2315 5X0 1 4631 0X0 2 9262 0X0 5 23155X1 46310X2 Ounce Troy 4 8240X0 0002 9 6480X0 0005 19 296X0 001 48 240X0 002 96 480X0 005 Penny weight 96 480X0 005 192 96X0 01 385 92X0 02 964 80X0 05 1929 6X0...

Page 13: ...TE 9600 bps DATA BIT 8 PARITY BIT N NONE STOP BIT 1 CODE ASCII BIT FORMAT LSB MSB 0 1 2 3 4 5 6 7 START BIT DATA BIT STOP BIT DATA FORMAT S U T G N W D1 D2 D3 D4 D5 D6 U1 U2 U3 U4 U5 LF CR SP SP WEIGHT SP UNIT ST STABLE UT UNSTABLE GW GROSS WEIGHT SP SPACE NW NET WEIGHT D1 D6 WEIGHT U1 U5 UNIT Examples ST GW 150 00 g UT NW 80 00 ct SCALE 褞砃DSUB9 5 9 4 8 3 7 2 6 1 GND TXD RXD BI DIRECTION DSUB9 ...

Page 14: ...Precision Balance 13 2007 5 22 ...

Reviews: