Precision Balance
- 6 -
Linearity Calibration of Weight
Step I
Power off while holding down
keys, and then press
「
ON/OFF
」
to switch on the scale.
Step II
Clear the weighing pan, then press
key, (zero point
calibration)
Step III Wait for a beep sound
Step IV
Place 1/3 of load on the weighing pan and press
to
perform 1/3 load calibration. Wait for a beep sound.
Step V
Place 2/3 of load on the weighing pan and press
to
perform 2/3 load calibration. Wait for a beep sound.
Step VI
Place full load on the weighing pan and press
to perform
full load calibration. Wait for a beep sound.
Step VII
Clear the weighing pan, then press
key
Step
VIII
After countdown, the linearity calibration is completed and the
scale returns to the normal weighing mode.