33
Technical specifications (EMC-RF)
Use only cables with the following specification:
USB shielded maximum 2 m
Serial shielded maximum 1.5 m
Guidance and manufacturer´s declaration — electromagnetic emissions
The HemoCue systems are intended for use in the electromagnetic environment specified below. The customer or the user of the HemoCue
systems should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment — guidance
RF emissions
Group 1
The HemoCue systems uses RF energy only for its
internal function. Therefore, its RF emissions are very
low and not likely to cause any interference in nearby
electronic equipment.
RF emissions
Class B
The HemoCue systems are suitable for use in all estab-
lishments, including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Complies
NOTE 1
It is the manufacturer’s responsibility to provide equipment electromagnetic compatibility information to the customer or user.
NOTE 2
It is the user’s responsibility to ensure that a compatible electromagnetic environment for the equipment can be maintained in
order that the device will perform as intended.
Summary of Contents for Plasma/Low Hb
Page 1: ...HemoCue Plasma Low Hb Operating manual Manual de funcionamiento Manuel d utilisation ...
Page 2: ......
Page 6: ...1 5 4 3 2 6 6 Components Componentes Composants ...
Page 8: ...1 2 3a 5 3c 4 3b 8 Start up Activación Démarrage ...
Page 10: ...5 3 4 2 2 1 10 Measuring Sample material Medición Tipo de muestra Mesure Échantillons ...
Page 12: ...7 8 6 9 12 ...
Page 14: ...2 1 14 Maintenance Mantenimiento Maintenance ...