9.1.3
Diagnostic messages
The reading out of the messages takes place together with the data and must be initiated by the
program of the master control system, see chapter 9.1.
Messages that can be read out over response data are:
Value valid: The "Value valid" bit is set to invalid as long as the following situation prevails:
- The AD converter switches to overflow.
- A calculation for determining the measurement value switches to overflow
- One or several sampling points violate the monotony.
Ready for operation
Bridge feed voltage is too small
Bridge feed voltage is too large
Wire break
Writing/reading error for offset/sampling point
ADC overflow or underflow
Control command executed
A configuration error is reported via the corresponding diagnostics.