Helmer Scientific i.Series
®
Platelet Incubator
Operation Manual
360388/A
5
2 Installation
2.1
Location Requirements
Note
Hot ambient temperatures with high humidity may cause condensation on the outside of the unit.
♦
Has a sturdy, level surface.
♦
Has a grounded outlet meeting national electric code (NEC) and local electrical requirements.
♦
Is clear of direct sunlight, high temperature sources, and heating and air conditioning vents.
♦ Meets limits specified for ambient temperature (15 °C to 35 °C) and relative humidity.
♦ Minimum 24” (610 mm) above for ambient temperatures of 28 °C to 35 °C.
♦ Minimum 4” (102 mm) above for ambient temperatures of 15 °C to 28 °C.
♦ Minimum 12” (305 mm) behind for ambient temperatures of 28 °C to 35 °C.
♦ Minimum 4” (102 mm) behind for ambient temperatures of 15 °C to 28 °C.
2.2
Placement and Leveling
Note
Rear stand-off brackets are provided with the unit and should be installed prior to placing the incubator in its location.
1.
Align keyhole openings in stand-off bracket with screws on back of
incubator and slide down to engage.
2. Tighten screws using a #2 Phillips screwdriver to secure.
3. Place platelet incubator on study surface.
4. Ensure platelet incubator is level.
2.3
Chart Recorder
Note
For complete information, refer to the Temperature Chart Recorder Operation and Service Manual.
A B F
C
D
E
Chart recorder with paper and battery installed.
Table 1. Chart Recorder
Label
Description
Function
A
Left and Right Arrow buttons Adjust settings and stylus position
B
LED
Indicates status of chart recorder in operating mode, or selected temperature range in paper change mode
C
Chart change button
Adjust position of stylus when changing chart paper, or run a test pattern
D
Stylus
Mark temperature line on paper
E
Reset button
Restart chart recorder
F
Backup battery
Provides power during AC power failure. Connect prior to use.
Rear Stand-off Bracket