![Helmer i.Series PC100-Pro Operation Manual Download Page 16](http://html.mh-extra.com/html/helmer/i-series-pc100-pro/i-series-pc100-pro_operation-manual_2122195016.webp)
Helmer Scientific i.Series
®
Platelet Incubator
Operation Manual
360388/A
15
6 Product Specifications
6.1
Operating Standards
These units are designed to operate under the following environmental conditions:
♦
Indoor use only
♦
Altitude (maximum): 2000 m
♦ Ambient temperature range: 15 °C to 35 °C
♦ Relative humidity (maximum for ambient temperature): 80% for temperatures up to 31 °C, decreasing linearly to 45% at 35 °C
♦ Temperature control range: 20 °C to 35 °C
♦
Overvoltage category: II
♦
Pollution degree: 2
♦
Sound level is less than 70 dB(A) under normal operation
♦
RF Emissions: Group 1 - Class A
♦
EMC Environment: Basic
Notes
•
Power Draw and Power Consumption specifications include internally operating Pro Series agitator supplied from the
Pro Series incubator through a 24 VDC umbilical cable (DC power cord).
• Power Draw measured in Watts.
• 100 V models have only 1 circuit breaker
Table 4. Electrical Specifications
PC100-Pro
PC900-Pro
PC1200-Pro
monitor battery backup (12V)
w/ system battery backup
Input Voltage and Frequency
100-240 VAC, 50 Hz / 60 Hz
Voltage Tolerance
±10%
Circuit Breakers
7.0A quantity 2
(100V units - quantity 1)
Agitator Power Draw (if installed)
16 W at 24 V (DC)
16 W at 24 V (DC)
16 W at 24 V (DC) 16 W at 24 V (DC)
Incubator Power Consumption
(with agitator installed)
65 Watts*
352 Watts**
75 Watts*
415 Watts**
65 Watts*
352 Watts**
65 Watts*
352 Watts**
Power Source
Varies (refer to product specification label)
Remote Alarm Capacity
1A at 33V (AC) RMS or 30V (DC)
Internal Outlet Maximum Current Draw
1.5A AT 24V (DC)
* 22°C in 24°C ambient static operation
** Pull down (full power)
CAUTION
• The interface on the remote alarm monitoring system is intended for connection to the end user’s central alarm
system(s) that uses normally-open or normally-closed dry contacts.
• If an external power supply exceeding 30 V (RMS) or 30 V (DC) is connected to the remote alarm monitoring system’s
circuit, the remote alarm will not function properly; may be damaged; or may result in injury to the user.
Note
Add 1.25” (31.75 mm) to depth of the PC100-Pro for handle.
Table 5. Incubator Specifications
100
900
1200
Height
27.2” (689 mm)
30.4” (772 mm)
30.4” (772 mm)
Width
20.8” (528 mm)
26.0” (661 mm)
40.3 (1023 mm)
Depth
22.3” (565 mm)
30.2” (766 mm)
30.2” (766 mm)
Weight
116 lbs (53 kg)
136 lbs (62 kg)
173 lbs (79 kg)