4.4. Advanced Parallel Operation with TCP/IP
Ethernet Connection
View the following settings and information on the Device Setup page
using a TCP/IP Ethernet connection to a unit: (see section 5)
Parallel Mode:
This field sets the Parallel Mode of the power supply
to Primary, Secondary, or Standalone. The Parallel Mode can also be
set from the Control Panel.
Device ID:
This field sets the unique Device ID (number between 0
and 5) for parallel operation of the power supply. The Device ID can
also be set from the Control Panel.
Send Secondary Alarm Emails:
If this box is checked while this
power supply is operating as Primary unit, the Primary will send out
alarm notification e-mails when an alarm occurs on a Secondary unit.
This feature is useful if your site has a limited number of Ethernet
network ports available and you only connect the Primary unit to the
Ethernet network.
•
This box will have no effect if the unit is configured as a
Secondary or Standalone unit
•
If this feature is enabled, check the Send E-mail boxes on the
Primary unit’s Alarm Setup page for the alarms you wish to
receive e-mail notifications for
Parallel Errors:
Displays any errors or alarm conditions related to
the serial SHARE interface that are currently active.