background image

- 44- 

Revision 4.6  3/2004

2 –

 The percent life remaining is based on the character-

istics of standard 9 Vdc alkaline batteries. Use of other 
batteries types, such as; zinc, lithium or rechargeable 
NiCad batteries, will result in inaccurate estimation of 
remaining battery life in both percent indication and bar 
graph indication. 

Step 1 

Install new batteries as outlined in the base unit 

start up section of this manual (Section 3.1).

Step 2 

Turn the HHC on.

Step 3  

 With the HHC operating and displaying pres-

sure press the battery check key, designated BAT CK:

 

BAT

 

CK

The HHC will respond by displaying:

 

+x.xxV    +xx%

Note:

 Installation of two new 9 Vdc alkaline batteries 

will produce an initial voltage measurement of 9.0-9.5 
Volts on the calibrator. If the value measured is signifi -
cantly below this level it is likely that the batteries have 
degraded in storage and that they will provide less than 
the 30 hours of operation expected from a set of fresh, 
fully charged batteries.

Step 4  

 Use the right arrow key to restore the bar graph 

to the full scale. The percentage value displayed will 
refl ect the adjustments made to the bar graph.

The number of highlighted segments in the bar graph 
will automatically reduce as the remaining battery life is 
reduced through the use of the HHC.

Important Note:

 It is recommended that batteries be re-

placed when the low voltage icon appears on the instru- 
the low battery icon appears is approximately 1 hour. 

****WARNING****

Batteries must be changed in a non-hazardous environ-
ment for all HHC units.

 Section 21.0   Dual Sensor Functions

The calibrator is equipped with fi rmware that can auto-
matically add the measured pressure values from two 
installed Quick Select pressure modules or to subtract 
the value measured by the Quick Select module installed 
in the right module bay from the measured value of the 
Quick Select value in the left module bay. This optional 
fi rmware allows for the following:

The ability to subtract the measured value from the left 
module of the measured value derived from the right 
module allows for simplifi ed measurement of what are 
commonly referred to as high-line” or elevated static 
differential pressures. A complete description of this 

capability is provided in Section 21.2 entitled Dual 
Sensor Differential Pressure Measurement.

2 –

 The ability to add the measured values of the two 

installed Quick Select pressure module allows for the 
addition of a measured barometric pressure from an 
absolute pressure module to the measured  pressure 
from a gauge or differential pressure. This capability can 
come in handy and save money by allowing the use of 
gauge measurement Quick Select pressure modules for 
absolute pressure measurement activities.

  Section 21.1   Accuracy of Dual Module 
 

 

Measurements

The accuracy of measurements made using the differ-
ence or sum of 2 modules is a function of the full scale 
range of the Quick Select pressure modules used and, 
in the case of differential pressure measurements, the 
magnitude of the dp range. 

For dual sensor summation measurements:

The worst case inaccuracy would be the sum of the 
inaccuracies of the two Quick Select modules used for 
the measurement. For example, if the measurement is 
made with two +0.1% full scale modules the maximum 
inaccuracy would be +0.2%. In actual use, it is gen-
erally acceptable to use the Root Sum of the Square 
(RSS) method of determining the accuracy of a multiple 
component measurement system. Using the RSS method 
the accuracy of the dual module summation is approxi-
0.14% of span.

For dual module differential pressure measurement:

The accuracy of the dual module differential pressure 
measurement is a function of the full scale range of the 
Quick Select pressure modules used in the measurement 
process and the magnitude of the dp range. In general, 
the accuracy for the dp measurement for dp ranges less 
than 10% of the rated range of the Quick select modules 
used is ±0.37% of dp range ±1 count. For dp ranges 
greater than 10% of the range of the Quick Select pressure 
module installed, the accuracy of the dp measurement is 
±0.37 % of the dp range x 0.1(R1) where R1 equals the 
ratio of the module full scale to the dp full scale (Sensor 
full scale/dp full scale). 

  Section 21.2 

Setting Up Dual Module Differential 

 

 

Pressure Measurement

Step 1  

 With the calibrator off, install the two Quick 

Select pressure modules in the calibrator. Be certain 
to install the module so that the modules are in the 
desired module bays. The subtraction process is always 
defi ned as the subtraction of the measured value of the 
right module from that of the measured value of the left 
module (d/p = Left measurement value – right measure-
ment value)

Summary of Contents for PTE-1

Page 1: ...PTE 1 HAND HELD CALIBRATOR OPERATING MANUAL REVISION 4 6 3 2004 I M002 10052 11 00 HHHC MAN 1P6 00 7P10 07 250 AMR...

Page 2: ...ur expert factory recalibration services please complete the material return form which can be printed directly from the service button on the Heise website Please return the instrument to Ashcroft In...

Page 3: ...perator Information 11 3 5 Removing Quick Select Modules 12 4 0 Key Function Overview 12 5 0 Port Select Function 13 6 0 Zero Function 14 6 1 Zeroing One of Two Installed Quick Select Pressure Module...

Page 4: ...imultaneous Display of Flow Velocity or Volume and An Electrical Output Measurement An Electrical Output Measurement An Electrical Output Measurement 28 15 0 Leak Detection Function 29 15 1 Leak Rate...

Page 5: ...ion Generation Enhanced Data Logging Mode 50 23 1 Set Up and Use of Certi cation Generation Data Logging Mode 50 24 0 HHC System Set Up for Uploading Stored Data Into an IBM Compatible PC 55 24 1 Inst...

Page 6: ...into either of the 2 module bays in the base unit Pressure engineering units can be independently selected from the library of twelve factory programmed or one operator designated engineering unit Te...

Page 7: ...tive strain gauge sensor This technology takes advantage of the fact that when put under stress as with the exing of a diaphragm under pressure or vacuum the resistive properties of a piece of silicon...

Page 8: ...interface module can accom modate up to 8 different programmed RTD calibration curves Programming of the RT module is accomplished via an optional software package and any PC compat ible computer wit...

Page 9: ...ure detector into the connector on the module To set up the HHC and temperature module combination proceed to section 16 for RTDs or section 17 for thermcouple detectors Important Note If only one mod...

Page 10: ...tional details on zeroing the HHC system are provided in Sec tion 6 0 of this manual The HHC is now ready for basic pressure or temperature measurement Simply connect the pressure port s of the Quick...

Page 11: ...timate inaccurate This function should be set up only when new batteries are installed If this function is not set up pressing the BATCK key will provide the measured remaining voltage of the installe...

Page 12: ...ion oriented functions the use of a real time clock will be required Setting the real time clock is a simple process and should be done when the HHC is received The real time clock may be set by follo...

Page 13: ...lear the entry and restart the access code entry process If the password codes have been lost contact your calibrator supplier The serial number of the base unit is required for the factory to provide...

Page 14: ...0 Key Function Overview The HHC has a great deal of functionality Many of the product capabilities are activated by dedicated keys on the keypad Other functions are either activated or set up through...

Page 15: ...ugh the set up functions provided by the corresponding menu options accessed and available by pressing the setup key Complete details of each of the above outlined capabili ties is included in the sec...

Page 16: ...have the following information on the display psi mA 12345 1234 Step 2 With the pressure display deactivated for the side that is not to be zeroed press the zero key ZERO The HHC will respond by disp...

Page 17: ...t abbreviations correspond to the following engineering units psi pounds per square inch inHg inches of mercury inH2O inches of water column ftSW feet of sea water Bar bar mBar millibar kpa kiloPascal...

Page 18: ...down arrow keys to increment or decrement through the alphabet numerical values and the left and right arrows to move the active entry eld cursor to the next loca tion If needed for the desired engin...

Page 19: ...either of the two desired engineering units For this example we will select psi as one of the desired units With psi selected the text psi will ash on the display and the display will look as follows...

Page 20: ...truments when calibrating pressure measurement devices in terms of the height of a column of water To facilitate your calibration and test of these instruments the HHC includes 3 conversion factors fo...

Page 21: ...lt value or to change the existing tare value If no tare value is desired input zero 0 Warning Failure to enter the zero 0 will result in the HHC automatically accepting the last displayed value as th...

Page 22: ...and available for transmission over the RS232 interface When damping is not activated the HHC takes a new pressure value every 130 to 240mS 8 times per second and updates the display If required this...

Page 23: ...f damp ing established in the set up menu is stored in EEPROM and therefore does not need to be reprogrammed after power up unless the level of damping desired has changed Section 13 0 Percent Functio...

Page 24: ...7 With the desired output zero keyed in and displayed press the enter key ENT The HHC will respond by displaying Enter output FS 0 0000 Step 8 Using the number keys enter the full scale output of the...

Page 25: ...ir density can further be calculated through employing the following equation dpw V 1097 x K x d d Adw Adw Ad 0 07649 x Pa 14 73 x 520 T 460 where Paa the absolute pressure in duct in psia T air tempe...

Page 26: ...it of measure for ow velocity follow the steps below Step 1 With the HHC operating in the ow velocity mode refer to Section 14 1 for instructions on entering the ow velocity mode press the engineering...

Page 27: ...ess the enter key ENT The HHC will respond by returning to the ow measure ment mode and will display the measurement data as follows ft min ft min 12345 12345 W width of the duct in inches H height of...

Page 28: ...min ft sec MPH MeterSec Step 7 With the text MeterSec selected and ashing press the enter key ENT The HHC will respond by displaying the following MeterSec mA 12345 1234 Step 8 Reactivate the right s...

Page 29: ...rement from a second installed Quick Select module For example the measurement of ow can be displayed in ft min for one installed Quick Select pressure module and the second installed pres sure module...

Page 30: ...he ow key on the keypad of the HHC FLOW The HHC will respond by displaying Flow Velocity Velocity Volume Step 4 Using the left right arrow keys select the desired measurement function In this case we...

Page 31: ...ibration of measurement instruments Refer to Section 13 0 for percent function set up and use information Section 15 0 Leak Detection Function The HHC provides the ability to detect and quantify leaks...

Page 32: ...ing the desired Quick Select pressure module must be monitored by the HHC system The following provides information on activating the desired pressure module for leak test applications With Two Quick...

Page 33: ...installed in the other sensor bay it can be re selected using the port select function PORT SEL as detailed in Section 5 0 Section 16 0 Temperature Measurement using RTD Probes The HHC can be used wi...

Page 34: ...ill be made in degrees Celsius Engnring units o C o F K o R Step 11 With the desired engineering unit selected and ashing press the enter key ENT The calibrator will respond with the following display...

Page 35: ...h the desired resolution selected and ash ing press the enter key ENT The calibrator will brie y show internal limit settings and will then commence temperature measurement The measurement parameters...

Page 36: ...ternal reference junction cannot be applied in this mode No further base unit programming will be required If the millivolt mode has been selected the programming cycle will end when the enter key is...

Page 37: ...mmed with the reference temperature of the external reference junction the display will read Ref temperature A value may already be programmed or the calibrator may display 0000 Entering a new value w...

Page 38: ...sh Disable Manual Deadband Step 3 With the word Deadband ashing press the enter key ENT Step 4 Gradually increase or decrease as required pressure vacuum or temperature until the display of the HHC in...

Page 39: ...Terminal To communicate with the HHC through the use of a dumb terminal set the terminal up as follows Select baud rate to be used 300 1200 2400 or 9600 Select communications com port 1 or 2 Set data...

Page 40: ...the measurement value displayed on the right side of the display including the selected unit of measure Section 19 3 Setup of the RS232 Interface in the Journal Mode To use the RS232 interface in the...

Page 41: ...ly suspend the transmis sion of data during the setup process For most functions data transmission will resume when the set up is com plete However the RS232 interface must be re enabled via the RS232...

Page 42: ...ll respond by displaying the following Inquiry Char xx x Step 12 Using the number keys enter the desired end of message character in decimal code A list of the avail able ASCII characters and their co...

Page 43: ...ter key a second time the HHC will respond by displaying the following LeftMod dd dd dd Type x where dd dd dd the date of the last calibration of the left module Type provides information on the type...

Page 44: ...1 2 entitled Dual Sensor Differential Pressure Measurement 2 The ability to add the measured values of the two installed Quick Select pressure module allows for the addition of a measured barometric p...

Page 45: ...PORT SELECT PORT SEL key First press of the port select key L R PSI PSI xxxx xxxx Second press of the port select key PSI L R PSI xxxx xxxx Third press of the port select key PSI mA xxxx xxxx Fourth p...

Page 46: ...me after the set up has been complete The calibrator will display the fol lowing text L R PSI PSI xxxx xxxx for a calibrator displaying measurement values in the psi engineering unit When the dual mod...

Page 47: ...d number the measured value and the engineering units of the measurement value If the time date stamp is enabled the time hours minutes seconds and date Year Month Day are included with the logged mea...

Page 48: ...label information is displayed press the enter key ENT The label is now stored and will appear in sequence with the corresponding data logged values when the stored information is viewed on the displ...

Page 49: ...e enter key to disable the data log function ENT Automatic time interval based data logging has now been terminated Section 22 5 Manual Data Logging Standard Data Logging Mode Follow the procedure bel...

Page 50: ...HHC system on and reading pressure press the data log key on the keypad DATA LOG The HHC system will respond by displaying the follow ing THE LAST DATA LOG MENU SCREEN USED Step 2 Using the up down an...

Page 51: ...lay THE LAST DATA LOG MENU SCREEN USED Step 2 Using the up down and right left arrows browse through the data log menu and locate the word Review The menu line with the word Review appears as Review O...

Page 52: ...data is facilitated by the enhanced data labeling capability provided by this mode of operation Section 23 1 Set Up and Use of Certi cation Generation Data Logging Mode Step 1 With the HHC on and disp...

Page 53: ...ion of a certi cation sheet press the Enter key with the text New Tag ashing ENT The HHC will respond by displaying the following Enter Tag Name Use the up and down arrow keys to enter letters and or...

Page 54: ...p 8 Using the numeral keys enter the desired number of as found data points to be logged In this example 10 data sets will be stored in the calibrator s data logging memory Enter of Testpts 10 Step 9...

Page 55: ...ter Step 14 To proceed with the entry recording of the ambient temperature value with the text Entry ashing press the enter key ENT Step 15 Note If the HHC has an GQS RT or TC temperature module and t...

Page 56: ...to be logged DATA STORE After the last set of data to be logged has been stored the HHC will prompt the operator with the following display AsLeft readings View Done Redo The options available are as...

Page 57: ...pproximate data logging memory available for storage of additional measurement and calibration data Step 28 Press any key to return to the rst CertGen Screen as follows New Tag Previous Tag Highlighti...

Page 58: ...s Follow the procedures below to change the battery back up unit Note The HHC system should be turned off prior to initiating battery replacement procedure Step 1 Turn the HHC indicator bottom side up...

Page 59: ...has been input and appears on the display press the enter key ENT The HHC system will respond by displaying HHMMSS Stop XXXXXX Step 6 Enter the desired stop time using the number keys When properly en...

Page 60: ...sable Enable In the above example the date time stamp function has been enabled and the stored data will be labeled with time date information in the database Step 9 With the desired format displayed...

Page 61: ...erical value of 10 The HHC will not automatically convert the entered trip point to a new value to re ect a change in engineering units Open Close If Open is selected the low limit contact will be ope...

Page 62: ...ritical to ensure the combined output of pressure module and base unit is within rated accuracy Section 27 2 Required Equipment The following equipment is required to perform recali bration or recerti...

Page 63: ...aution when connecting test leads and applying test voltage inputs The Model CQS Calibration Quick Select module is available to simplify connection of the voltage stan dard to the base unit The modul...

Page 64: ...RTmodule Step 4 With the text Calib displayed and ashing press the enter key ENT Step 5 The calibrator will respond by displaying Access code 000000 Step 6 Enter the 5 digit calibration access code pr...

Page 65: ...volts applied press the enter key ENT The calibrator will respond by displaying xxxxx working After the calibrator completes auto adjusting the mea sured value it will display Apply 10 0000V to volts...

Page 66: ...andards must be used to assure the accuracy of the calibrator system is not degraded by the calibra tion process As recommended by standards organiza tions such as A S M E N I S T and various U S and...

Page 67: ...librator will dis play Full Scale Cal Yes No Step 14 To perform full scale calibration adjustment select Yes and press the enter key ENT To skip full scale adjustment and proceed to re lineariza tion...

Page 68: ...nearize a Quick Select pressure module select the word Yes to the linearity query Important Note If full recalibration of the pressure module is desired the zero and span adjustment must be reset thro...

Page 69: ...ibration points needed to assure adequate correction for non linearity is as follows Sensor Accuracy Number of Points 0 025 0 05 20 including temperature 5 increments 0 025 0 07 11 10 increments and z...

Page 70: ...559 028959 1 001 383292 0319410 0097355 14 2233 980 665 0 980665 0 967842 98060 5 98 0665 1001 8 394 41 32 868 735 559 28 959 1000 1 383 292 31 9410 9 7355 255 851 255851 2 6136 1 02900 08575 1 91908...

Page 71: ...41 00101001 29 42 00101010 2A 43 00101011 2B 44 00101100 2C 45 00101101 2D 46 00101110 2E 47 00101111 2F 48 00110000 30 0 49 00110001 31 1 50 00110010 32 2 51 00110011 33 3 52 00110100 34 4 53 001101...

Page 72: ...93 148 10010100 94 149 10010101 95 150 10010110 96 151 10010111 97 152 10011000 98 153 10011001 99 154 10011010 9A 155 10011011 9B 156 10011100 9C 157 10011101 9D 158 10011110 9E 159 10011111 9F 160...

Page 73: ...0110 D6 215 11010111 D7 216 11011000 D8 217 11011001 D9 218 11011010 DA 219 11011011 DB 220 11011100 DC 221 11011101 DD 222 11011110 DE Hexa ASCII Decimal Binary decimal Character 176 10110000 B0 177...

Page 74: ...1 in H2O Sensitivity 0 002 of span typical Media Compatibility Clean dry non conductive non corrosive gas Under Overpressure Capability 15 to 50 psi Maximum Static line Pressure 100 psi Process Connec...

Page 75: ...bove are accessible through the unit select feature However readout will default to the pri mary unit of measure on start up Sensor modules scaled in primary units other than in H2 O GQS 1 or psi GQS...

Page 76: ...0 to 999 3 0 3 0 0 08 1000 to 1820 1 2 1 2 0 04 Max additional error Accuracy 25 C Accuracy 25 C due to ambient Thermocouple Measurement Range Ambient Not Including Ambient Including temperature devia...

Page 77: ...PPENDIX D RTD PROBE CONNECTOR SWITCHCRAFT TINI QG TA4F 2 WIRE RTD TA4F 1 2 3 4 3 WIRE RTD TA4F 1 2 3 4 4 WIRE RTD TA4F 1 2 3 4 PIN FUNCTION 1 2W 2 4W 3 4W 4 2W PIN FUNCTION 1 2W 2 4W 3 4W 4 2W PIN FUN...

Page 78: ...2 Attach copper wire to the and terminal posts of a precision millivolt calibrator source rated at 0025 or better 3 Wire the opposite end of the wires to a miniature U type copper copper thermocouple...

Page 79: ...79 Revision 4 6 3 2004...

Page 80: ...ice All sales subject to standard terms and conditions Ashcroft Inc Visit our web site www heise com Ashcroft Inc 250 E Main Street Stratford CT 06614 5145 Tel 203 378 8281 Fax 203 381 9042 email info...

Reviews: