Heinrichs Messtechnik GmbH
Ex- Supplement operating manual
UMC4 with TM/TMU/TME/TMR/TM-SH
Page
13 of 31
Temperature sensor circuit (terminals 5 to 8)
Voltage
Ui DC
30 V
Current
Ii
100 mA
Power
Pi
0,1 W
Effective internal capacitance
Ci
negligible
Effective internal inductance
Li
negligible
4.4.2
Ambient temperature range Ta
Depending on type of connection, installation, process temperature and temperature class:
With plug
Adapter
Process temperature
(°C) (1)
Ambient temperature range
(°C) (1)
Temperature class
without
40
-40 to +40
T6
without
60
-40 to +60
T5
without
60
-40 to +10
T4
60 mm
100
-40 to +100
T4
160 mm
120
-40 to +100
T4
160 mm
180
-40 to +100
T3
260 mm
220/260
(2) (3)
-40 to +100
T2
(1) The maximum ambient and process temperatures of the TM-SH are restricted to 60 °C and 100 °C respectively
(2) The maximum process temperature of 260 °C only short-term. Refer to section 3.1 for more details.
(3) The process temperatures 220/260 °C are only applicable for TM, TMU and TMR sensors
Remote mount configuration
Adapter
Process temperature
(°C)
Ambient temperature range
(°C)
Temperature class
without
40
-40 to +40
T6
without
60
-40 to +60
T5
without
100
-40 to +100
T4
100 mm
120
-40 to +100
T4
100 mm
180
-40 to +100
T3
200 mm
220/260
(2)
-40 to +100
T2
(2) The maximum process temperature of 260 °C only short-term. Refer to section 3.1 for more details.
Integral mount configuration
See chapter 4.5.4.5