Assembly
76
Condenser:
Glassware set G5 dry ice cooler
✓
The chiller is mounted on the drive.
➜
Screw the ventilation cap onto the
opening opposite the vapor tube.
➜
If you are working with inert gas,
screw the inlet hose with inert gas
valve onto the opening above the
ventilation cap.
Inert gas valve
Ventilation cap
➜
Place the red seal ring in the groove at
the top of the chiller flange.
➜
Place the white PTFE ring above it with
the collar facing upwards in the chiller
flange.
➜
Now insert the cold trap into the
condenser.
✓
The flange is drawn in by the vacuum
and the system is closed vacuum-tight
by the installed seal.
➜
Place the black lid on top to protect
your dry ice.