Troubleshooting
61
After the rotary evaporator has been switched on, it checks the attached components.
The scope of functions depends on the devices attached.
Missing
component
Description /
Response of the device
Remedy
Control box
Control box is not
attached:
▪
Vacuum value is not
shown in display
▪
Vacuum control deacti-
vated
▪
Attach control box and components
▪
Switch the device off and then on
again
Vacuum valve or
vacuum pump
Vacuum valve or vacuum
pump is not attached:
▪
Vacuum control is
deactivated
▪
Attach components
▪
Switch the device off and then on
again
Vapor temperature
sensor:
The sensor for deter-
mining the vapor
temperature is not
attached:
▪
Steam temperature is
not displayed
▪
Attach components
▪
Switch the device off and then on
again
AUTOaccurate
sensor
The sensor for controlling
the automatic mode DAA
is not attached:
▪
“Automatic (DAA)”
mode is deactivated
▪
Attach components
▪
Switch the device off and then on
again