![Heidolph 036308001 Operating Instructions Manual Download Page 140](http://html1.mh-extra.com/html/heidolph/036308001/036308001_operating-instructions-manual_2118931140.webp)
140
Fault
`
Possible cause
3
Remedy
Personnel
Sensors
submenu is per-
manently displayed
`
Submenu
Sensors...
does not automatically
switch back to previous
display.
3
Select the required sensor
by turning and pressing
selection knob.
User,
Specialist
After loading defaults
Language selection
appears
`
Special factory settings
have been loaded.
3
Set language and pressure
unit.
IMPORTANT!
Check if the
loaded default settings are
suitable for your vacuum ap-
paratus.
respon.
Specialist
Error I/O module
`
Plug disconnected.
`
An error occured in the
system, the I/O module
passed the error alert to
the controller.
3
Check plug connection.
3
Remedy external fault.
Specialist,
respon.
Specialist
VSP sensor displays
wrong values
`
VSP sensor configured
as VSK.
3
Use menu
Function/
Vacuubus
to reconfigure
the sensor as VSP.
Specialist,
respon.
Specialist
Controller in operation,
pressure display flashes
`
VSK sensors are measu-
ring negative difference
pressure.
3
Select an other vacuum
sensor in menu
Sensors
.
Action required
`
Cause
3
Remedy
Elapsed process time
`
All program steps are
completed.
`
Program end reached.
3
Acknowledge indication by pressing
Start/Stop
key.
Flashing clock icon
`
Elapsed process time
3
Acknowledge indication by pressing
Start/Stop
key.
Pump down stops,
flashing arrow down icon
`
Pressure below preset
minimum value.
3
Acknowledge indication by pressing
Start/Stop
key.
3
If possible readjust presetting (Min.).
Vac control stops,
flashing arrow up icon
`
Preset maximum value
exceeded.
3
Acknowledge indication by pressing
Start/Stop
key.
3
If possible readjust presetting (Max.).
Program -
`
Program not yet stored.
3
Store program under a free program
number.
Summary of Contents for 036308001
Page 1: ...Betriebsanleitung Operating instructions Hei VAC Control Intelligent evaporation...
Page 6: ...Erste Schritte First steps Neuger t Erste Schritte Neuger t First steps delivery status...
Page 39: ...38 6 2 Sprache und Druckeinheit einstellen Sprache und Druckeinheit einstellen 5 Startbild...
Page 81: ...81 First steps delivery status First steps on delivery status Select language and units...
Page 114: ...114 6 2 Select language and pressure unit Select language and pressure unit 5 Initial screen...
Page 154: ......