![Heidolph 036308001 Operating Instructions Manual Download Page 136](http://html1.mh-extra.com/html/heidolph/036308001/036308001_operating-instructions-manual_2118931136.webp)
136
Combinations of flashing display icons
Icon flash rate
Fault and Meaning
beep
when Sound
On
On
`
Limit pressure reached
1x
`
Overpressure
1x
`
Process time elapsed
1x
+
`
Venting valve
2x
+
`
In-line suction valve
3x
+
`
Coolant valve
4x
`
External sensor removed
`
or defective
5x
`
internal sensor defective
7x
+
`
Vario pump
6x
+
`
VACUULAN process pressure not
reached within 99 hours.
8x
Digital I/O module:
`
Fault indicator triggered or
`
fault special configurations
9x
+
`
Level sensor triggered; flask full
10x
+
`
Emission condenser Peltronic (too
hot)
11x
`
Analog I/O module
12x
A defective I/O module, which is configured as a remote
module, does not trigger a warning alert. The control is
stopped. Alert display by the flashing warning triangle.