Argon Pressure Reducer for HBS Xtouch
12
EN
2 Warning
3 Introduction
By purchasing this instrument, you are now the owner of a high-quality product whose use and
correct handling are described in the following. In order to minimize possible risks to patients and
users, please observe these instructions carefully. Use, disinfection, cleaning and sterilization
may only be performed by suitably trained specialist personnel.
4 Scope
This product is designed to power HBS-Xtouch devices with argon gas. It reduces the gas
pressure of the gas source to the required operating pressure and maintains it constant.
4.1 Inteded Use
Argon pressure reducer: This product is designed to power Argon gas HBS x-touch devices. It
reduces the incoming gas pressure of the gas source to the required operating pressure and
maintains it constant.
4.2 Contraindication
No contraindications are known.
!
Incorrect use of the pressure reducer can cause serious damage.
!
Run the pressure reducer only at the intended pressure and with the intended gas. The use
of improper gas or at improper pressure might cause risks of explosion.
!
Protect the pressure reducer from dust, oil and other sources of dirt.
!
Do not use the pressure reducer if it is not in prefect working conditions.
!
When you draw gas, the cylinder must be placed upright and protected from falling.
!
Danger of explosion. At startup and also during use, hands and tools must be free from oil
and grease.