Trouble Shooting

Trouble

 

Resolution 

No power  
present

—  Verify that the AC adapter is fully plugged into 

the wall and rear of unit

—  Verify that the power switch is turned on

Unit or display 
is not operating 
normally

—  Turn off the unit, wait 2 minutes, and turn power 

on

Excessive 
vibration
Excessive noise

—  Inspect the tubes for equal fill or improper  

placement

—  Inspect the rotor for improper installation
—  Remove the tubes and determine if the noise  

persists with an empty rotor

Lid will not 
close

—  Verify that nothing is blocking the lid from fully 

closing 

—  Verify that nothing has fallen into the lock  

mechanism opening

Lid will not 
open

—  If for any reason you need to manually open the 

lid to access tubes due to an error or power loss, 
please perform the below

—  Turn off the unit, remove power cord
—  Make sure the rotor has stopped completely
—  Use a thin rod and insert it into the opening 

on the bottom  

—  Press gently but firmly.  You will feel a 
—  mechanical movement within the unit and 

the lid will release.

—  Remove the rod, set the unit on the feet
—  Remove your tubes and reclose the lid

Balancing the Rotor

Tubes of equal weight and size should be placed opposite each other.  
Use additional sample or water in other tubes to provide a balanced 
rotor.  Examples of proper sample balancing are illustrated next.

Warning!

Spin balanced loads only!
Do not place the tubes asymmetrically nor load unequal volume 
tubes.  Improper placement of tubes will lead to insufficient  
centrifugation and may cause serious injury or an accident.

Quick Spin Operation:

1.  Prepare the unit for centrifugation as in the General Operation  

(section above Steps 1-8).

2.  Press Quick Spin button (rotor should spin up to the set speed).
3.  The timer will begin counting.
4.  Once enough time has elapsed for your needs, release the Quick 

Spin button.  The rotor will spin down, stop and the unit will emit a 
beep when the cycle is complete.

5.  To open lid, press the OPEN button.

Other:  
Change Between RPM and RCF 

 

Prior to a spin cycle, press both the Speed Up & Down buttons 
simultaneously.  

The speed setting will then convert to the other.

Quick Stop

If for any reason you need to stop the cycle quickly, 

press and hold

 

the Start/Stop button.  The cycle will stop more rapidly than normal.   

Caution: 

 The tube contents may be disturbed during the quick 

stop process.

• 

If an error happens, the unit will beep and the display will indicate 
the error.

• 

Press Start/Stop to clear the error from the display.

Error:

Error Status

Resolution

Motor  
Overload

—  Something is interfering with the rotor.  Clear the rotor and 

reset.

User Stop

—  The operator has held down the Start/Stop and  

implemented a quick stop.

Balance

—  Inspect the tubes for equal tube fill or improper placement.  

Once corrected, rerun.

—  If the balance error continues to happen, remove the tubes 

and determine if the balance error still persists with an 
empty rotor.  

—  Inspect the rotor for improper installation. 

Temperature

—  The unit has exceeded the normal operating temperature.
—  Turn off the unit and allow to cool.

Excessive Tilt

—  The unit has experienced a non-normal tilt event.  
—  Make sure the unit is placed on a level surface.   

Once corrected, rerun

Lid Fail

—  The lid has opened during the cycle.  Check for proper 

operation of the lid lock mechanism.  The lid should stay 
locked during the entire cycle.

Rotor Lock

—  The unit has experienced a problem with the rotor.  
—  Correct the rotor interference.  Once corrected, rerun.

STATUS         ERROR
Rotor Lock

Status of Unit

Error Type

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for Gusto

Page 1: ...g x All rights reserved Printed in China Tous droits réservés Imprimé en Chine Todos los derechos reservados Impreso en Chine Alle Rechte vorbehalten Gedruckt in China Instruction Manual 26 015 10050 51 483 Heathrow Scientific LLC High Speed Mini Centrifuge www com information itm com 1 800 561 8187 ...

Page 2: ... manual contains important information safeguards and operating instructions 17 Follow all precautions from the Material Safety Data Sheet MSDS for any reagent you use with this instrument 18 Do not use solvent or flammable liquids near this or other electrical equipment 19 Do not centrifuge flammable explosive or corrosive materials 20 Do not operate in a hazardous or flammable environment 21 Do ...

Page 3: ... which case the user will be required to correct the interference Supplied Equipment 1 ea Centrifuge 1 ea Universal AC power adapter with 4 mains cables 1 ea Standard Tube rotor 1 5mL 2 0mL tubes 12 500 rpm 1 ea Strip tube rotor 8 place 0 2ml Strip Tubes or 32 individual 0 2ml tubes 12 500 rpm 12 ea 0 2mL tube adapters for use in the standard rotor only 12 ea 0 5mL tube adapters for use in the sta...

Page 4: ...flow An overheating situation could occur Verify that the power switch in the back of the unit is in the OFF position Plug the DC cord into the rear of the unit and the AC cord an approved outlet Display Display Status LID OPEN Lid is open READY Unit ready to centrifuge SPIN UP Cycle is starting and is increasing speed SPINNING Running in Standard mode SPIN DOWN Cycle is slowing down STOPPED Cycle...

Page 5: ...k Spin button rotor should spin up to the set speed 3 The timer will begin counting 4 Once enough time has elapsed for your needs release the Quick Spin button The rotor will spin down stop and the unit will emit a beep when the cycle is complete 5 To open lid press the OPEN button Other Change Between RPM and RCF Prior to a spin cycle press both the Speed Up Down buttons simultaneously The speed ...

Page 6: ...2 Strip 2 Tubes 4 Tubes 6 Tubes 8 Tubes 10 Tubes 12 Tubes Warning Do not attempt to open lid until rotor has completely stopped Avoid severe personal injury or property damage from moving parts Only use rotor compatible with the centrifuge Warning Do not use damaged or cracked rotors Rotor installation and removal To remove rotor With one hand grasp the rotor With the other hand unscrew center kno...

Page 7: ...LIGATION UNDER THIS WARRANTY EXCEED THE PRICE OF THE PRODUCT Heathrow Scientific LLC will not accept any product return that is not appropriately cleaned and decontaminated In the unlikely event of repair or when damage to the unit necessitates return contact Heathrow Scientific LLC and obtain return authorization before sending your product for service Return for Repair Important Transporting haz...

Reviews: