Your mini-centrifuge is designed for laboratories in the bioscience,  
medical, and chemistry fields.  It may be involved with biohazardous 
and/or hazardous materials.  This instruction manual cannot address all 
safety hazards.  It is the responsibility of the user to consult and observe 
all health and safety precautions and to assess the instrument’s  
suitability to the task.

1.  Never use this product in any manner inconsistent with these  

instructions.

2.  This product is intended to be used indoors only.
3.  This product is intended for separating aqueous solutions in 

approved test tubes compatible to the RCF of this unit.

4.  The product is not intended to centrifuge very dense materials.  Do 

not use materials with a density greater than 1.2g/cm.

5.  Observe general laboratory safety precautions and regulations when 

using this product.

6.  Do not attempt to operate with the cover removed or open.
7.  The rotor and lid must always be securely fastened when in  

operation.

8.  Always seat the rotor completely onto the shaft, then tighten the rotor 

nut.

9.  Do not use damaged rotors.
10.  Do not put hands onto the rotor unless the rotor is completely 

stopped.

11.  Do not move unit while the rotor is spinning.
12.  Do not immerse the product in water.  Injury due to shock or fire may 

result.

13.  The rotor must be loaded symmetrically.  Operating this unit with an 

unbalanced rotor will cause severe vibration and damage, which is 
not covered by warranty.

14.  Do not fill tubes while inserted in rotor.  Spilled liquid may harm unit.
15.  Close tube lids before starting centrifuge.  Open lids can be torn off 

during centrifugation and damage the unit.  Open lids can cause fluid 
to be dispersed inside the centrifuge.

16.  Do not insert non-standard tubes or other foreign articles into the 

rotor.

Safety Precautions

Caution!

Before starting the mini-centrifuge for the first time, please read 
this instruction manual carefully.  This manual contains  
important information, safeguards, and operating instructions.

17.  Follow all precautions from the Material Safety Data Sheet (MSDS) for 

any reagent you use with this instrument.

18.  Do not use solvent or flammable liquids near this or other electrical 

equipment.

19.  Do not centrifuge flammable, explosive, or corrosive materials.
20.  Do not operate in a hazardous or flammable environment.
21.  Do not use any solvent on the unit that may attack plastic or cause 

cracks in the rotor.

22.  Ensure the rotor is protected from corrosion and mechanical damage. 

The rotor must be cleaned with a pH-neutral cleaning liquid.

23.  Always work in a manner which endangers neither the user nor any 

other person.

24.  Should the instrument fail to work to specification, immediately stop 

centrifuging.  Clean and troubleshoot the instrument according to the 
instructions under “Trouble Shooting” before any further use of the 
instrument.  Contact an authorized dealer or the manufacturer if trou-
ble persists.

25.  Do not attempt to stop rotor while unit is running.  Doing so may 

cause the unit to fail and will void the warranty.

26.  Only use original manufacturer’s rotors, AC adapters, tube adapters 

and any other spare parts.

27.  Repairs are to be performed by trained and authorized service per-

sonnel only.  Contact Heathrow Scientific

®

 LLC.

28.  Opening the instrument housing or improper use of the instrument 

voids the warranty.  If there is a failure during the warranty period, 
contact Heathrow Scientific® LLC for warranty service.

29.  Do not attempt to use units that have not been correctly installed or 

repaired.

30.  Do not attempt to disassemble or modify this product.
31.  Store at room temperature in a dry area.  Do not expose to sunlight, 

moisture, or extreme temperatures for prolonged periods of time.

Warning!

If the unit is not used as recommended by the manufacturer, the 
overall safety will be impared.

Warning!

Check MSDS, wear required Personal Protective Equipment, and 
observe all applicable local and national regulations before  
dispensing and disposing of hasardous sample.

www.

.com

[email protected]

1.800.561.8187

Summary of Contents for Gusto

Page 1: ...g x All rights reserved Printed in China Tous droits réservés Imprimé en Chine Todos los derechos reservados Impreso en Chine Alle Rechte vorbehalten Gedruckt in China Instruction Manual 26 015 10050 51 483 Heathrow Scientific LLC High Speed Mini Centrifuge www com information itm com 1 800 561 8187 ...

Page 2: ... manual contains important information safeguards and operating instructions 17 Follow all precautions from the Material Safety Data Sheet MSDS for any reagent you use with this instrument 18 Do not use solvent or flammable liquids near this or other electrical equipment 19 Do not centrifuge flammable explosive or corrosive materials 20 Do not operate in a hazardous or flammable environment 21 Do ...

Page 3: ... which case the user will be required to correct the interference Supplied Equipment 1 ea Centrifuge 1 ea Universal AC power adapter with 4 mains cables 1 ea Standard Tube rotor 1 5mL 2 0mL tubes 12 500 rpm 1 ea Strip tube rotor 8 place 0 2ml Strip Tubes or 32 individual 0 2ml tubes 12 500 rpm 12 ea 0 2mL tube adapters for use in the standard rotor only 12 ea 0 5mL tube adapters for use in the sta...

Page 4: ...flow An overheating situation could occur Verify that the power switch in the back of the unit is in the OFF position Plug the DC cord into the rear of the unit and the AC cord an approved outlet Display Display Status LID OPEN Lid is open READY Unit ready to centrifuge SPIN UP Cycle is starting and is increasing speed SPINNING Running in Standard mode SPIN DOWN Cycle is slowing down STOPPED Cycle...

Page 5: ...k Spin button rotor should spin up to the set speed 3 The timer will begin counting 4 Once enough time has elapsed for your needs release the Quick Spin button The rotor will spin down stop and the unit will emit a beep when the cycle is complete 5 To open lid press the OPEN button Other Change Between RPM and RCF Prior to a spin cycle press both the Speed Up Down buttons simultaneously The speed ...

Page 6: ...2 Strip 2 Tubes 4 Tubes 6 Tubes 8 Tubes 10 Tubes 12 Tubes Warning Do not attempt to open lid until rotor has completely stopped Avoid severe personal injury or property damage from moving parts Only use rotor compatible with the centrifuge Warning Do not use damaged or cracked rotors Rotor installation and removal To remove rotor With one hand grasp the rotor With the other hand unscrew center kno...

Page 7: ...LIGATION UNDER THIS WARRANTY EXCEED THE PRICE OF THE PRODUCT Heathrow Scientific LLC will not accept any product return that is not appropriately cleaned and decontaminated In the unlikely event of repair or when damage to the unit necessitates return contact Heathrow Scientific LLC and obtain return authorization before sending your product for service Return for Repair Important Transporting haz...