Page 30 of 36
101515-0 RMLD MANUAL REV
F
•
Low battery
recharge if necessary
Excessive false detections
while scanning at longer
distances
•
Scanning too fast
•
Alarm detection
threshold set too low
•
Scanning at the range
limit of the instrument
•
Slow down the scanning
rate. Pause at the long
range and sweep
towards you
•
Increase the Alarm
detection threshold
•
Move in closer
Excessive false detections
while scanning at closer
distances
•
Scanning too fast
•
Alarm detection
threshold set too low
•
Avoid making abrupt
motions while scanning
•
Increase the Alarm
Detection Threshold
Excessive false detections
or loss of sensitivity
•
Laser output not
optimized
•
Alarm Detection
Threshold set too high
or low for conditions
•
Perform Self-Test
procedure
•
Check the Alarm
Detection Threshold
Error icon or Warning icon
on continuously
•
Low battery
•
Moisture condensation
on mirror due to rapid
change in temperature
•
Internal component
failure
•
Check battery level and
recharge if necessary
•
Allow for the
temperature of the
instrument to stabilize
•
Note error code and
contact
HEATH
Battery indicator does not
show full charge after
charging
•
Battery not fully
charged
•
Battery level calibration
error
•
Charge unit until solid
green light on charger is
on
•
Run unit until it shuts
off, then fully recharge
with out interruption
Battery indicator bars are
flashing on and off
•
Battery level calibration
error
•
Run unit until it shuts
off, then fully recharge
with out interruption