Table 3 - Continued
For all ME EQUIPMENT and ME SYSTEMS
Radiated RF
IEC
61000-4-3
3 V/m
80 MHz
to 2.5 GHz
3 V/m
Recommended separation distance:
d=1.2
P
80 MHz to 800 MHz
d=1.3
P
800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to
the transmitter manufacturer
and d is the recommended
separation distance in meters
(m). Field strengths from fixed RF
transmitters, as determined by
an electromagnetic site survey, a
should be less than the compliance
level b in each frequency range.
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects and people.