Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF
communications equipment and the 04-630-001, 04-630-006, 04-631-001
and 04-631-006.
The 04-630-001, 04-630-006, 04-631-001 and 04-631-006 are intended for
use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the 04-630-001, 04-630-006, 04-631-001
or 04-631-006 can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment
(transmitters) and the 04-630-001, 04-630-006, 04-631-001 or 04-631-006
as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter W
Separation distance according to frequency of transmitter.
150 kHz to 80
MHz
d=1.2
P
80 MHz to 800
MHz
d=1.2
P
800 MHz to 2.5
GHz
d=1.2
P
0,01
0.12
0.12
0.23
0,1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23