Table 4

For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING

Recommended separation distances between portable and mobile RF 
communications equipment and the 04-630-001, 04-630-006, 04-631-001 
and 04-631-006.

The 04-630-001, 04-630-006, 04-631-001 and 04-631-006 are intended for 
use in an electromagnetic environment in which radiated RF disturbances are 
controlled. The customer or the user of the 04-630-001, 04-630-006, 04-631-001 
or 04-631-006 can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile RF communications equipment 
(transmitters) and the 04-630-001, 04-630-006, 04-631-001 or 04-631-006 
as recommended below, according to the maximum output power of the 
communications equipment.

Rated maximum 

output power of 

transmitter W

Separation distance according to frequency of transmitter.

150 kHz to 80 

MHz

d=1.2

 

P

80 MHz to 800 

MHz

d=1.2

 

P

800 MHz to 2.5 

GHz

d=1.2

 

P

0,01

0.12

0.12

0.23

0,1

0.38

0.38

0.73

1

1.2

1.2

2.3

10

3.8

3.8

7.3

100

12

12

23

Summary of Contents for 04-630-001

Page 1: ...SELECT DIGITAL BLOOD PRESSURE MONITOR Please read this guidebook completely before operating the unit Item 04 630 001 Item 04 631 001 Item 04 630 006 Item 04 631 006 English Spanish INSTRUCTION MANUAL...

Page 2: ...The purchaser must return the original Blood Pressure Monitor prior to a replacement product being provided Replacement of the product is the only remedy covered by the Limited 5 Year Warranty This w...

Page 3: ...use your monitor will provide you with many years of reliable readings To achieve the maximum benefit from your blood pressure monitor we recommend that you first consult with your physician or traine...

Page 4: ...p the inflation or deflation process anytime by pressing the POWER button PRODUCT IDENTIFICATION SYS DIA mmHg PUL min mmHg Arm Cuff AC Adapter Port Digital Display POWER Button on off MEMORY Recall Bu...

Page 5: ...ood pressure and display both systolic and diastolic readings on the screen Your systolic will be positioned as the upper number and the diastolic reading will be the lower number systolic diastolic 1...

Page 6: ...ition Because of this it is important to take consistent daily measurements of your blood pressure The variations in your individual readings should only be interpreted by your physician or a trained...

Page 7: ...et place You should be seated and relaxed 3 Avoid smoking eating taking medication alcohol consumption or physical activity 30 minutes prior to measuring your blood pressure If you are exhibiting sign...

Page 8: ...ic pressure 180 mmHg or Diastolic pressure 110 mmHg Stage 2 Hypertension Systolic pressure 160 179 mmHg or Diastolic pressure 100 109 mmHg Stage 1 Hypertension Systolic pressure 140 159 mmHg or Diasto...

Page 9: ...E Please consult with your physician or trained healthcare professional for further information regarding an irregular heartbeat if this symbol appears frequently at the end of a reading If Irregular...

Page 10: ...local regulations at the end of their usage NOTE Replace the batteries when the Low Battery symbol appears on the display or when the display remains blank when the POWER button is pressed 4 It is re...

Page 11: ...y 4 The date Fig 2 followed by the hour and minutes will flash Repeat steps 2 and 3 to adjust the date time setting 5 After setting the minutes the unit will automatically exit out of the date time se...

Page 12: ...or open the cuff by pulling or rolling the bottom of the cuff towards the right Fig 2 This should open the cuff creating a cylinder Do not extend the cuff beyond the metal bar 5 Insert your arm into...

Page 13: ...n Fig 5 6 When the display appears as shown Fig 6 the monitor is ready The cuff will automatically begin to inflate 7 Once the cuff inflates to the appropriate pressure level the measurement will begi...

Page 14: ...ton again to scroll through the saved measurements Fig 1 After you ve scrolled through all the readings the unit will automatically turn off NOTE Each memory bank can store up to 60 readings When the...

Page 15: ...unstable or it has too much air left in it Once the arrow stops flashing the unit is ready to take a measurement Appears when recalling a measurement from memory DIA SYS DIA SYS DIA SYS DESCRIPTION OF...

Page 16: ...plying Your Blood Pressure Cuff and Taking Your Blood Pressure Reading sections Replace the batteries EE symbol displays Do not talk or move during the measurement Review Applying Your Blood Pressure...

Page 17: ...with Ethyl alcohol 75 90 then dry the cuff by airing 9 No component of the monitor should be serviced by the user The circuit diagrams component part lists descriptions calibration instructions or oth...

Page 18: ...n Automatic Pressure Release Valve Manual fast release Memory Built in memory enabling display of up to 120 measurements Automatic shut off Approximately 1 minute after last button operation Operation...

Page 19: ...630 001 04 630 006 04 631 001 and 04 631 006 should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance RF emissions CISPR 11 Group 1 The 04 6...

Page 20: ...601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic til...

Page 21: ...006 04 631 001 or 04 631 006 should assure that it is used in such an environment IMMUNITY test IEC 60601 test Compliance level Electromagnetic environment guidance Radiated RF IEC 61000 4 3 3 V m 80...

Page 22: ...is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level b in each frequ...

Page 23: ...d If the measured field strength in the location in which the 04 630 001 04 630 006 04 631 001 04 631 006 are used exceeds the applicable RF compliance level above the 04 630 001 04 630 006 04 631 001...

Page 24: ...04 630 001 04 630 006 04 631 001 or 04 631 006 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and...

Page 25: ...requency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for...

Page 26: ...SAMPLE BLOOD PRESSURE LOG...

Page 27: ...STRUCCIONES MONITOR DIGITAL DE PRESI N ARTERIAL SELECT S rvase leer toda esta gu a antes de operar la unidad Art culo n 04 630 001 Art culo n 04 631 001 Art culo n 04 630 006 Art culo n 04 631 006 Ing...

Page 28: ...esi n Arterial original antes de que se proporcione un producto de reemplazo El reemplazo del producto es el nico recurso que ofrece la Garant a limitada por 5 a os Esta garant a le otorga derechos le...

Page 29: ...n sangu nea Con el uso y cuidado adecuados su monitor le proporcionar muchos a os de lecturas confiables Para obtener el mayor beneficio de su monitor de presi n arterial le recomendamos que primero...

Page 30: ...Hg Brazalete para el brazo Puerto para adaptador AC Conector del tubo Visor Digital Bot n para llamar la MEMORIA Bot n de encendido encendido apagado PRECAUCI N Puede detener el proceso de inflado o d...

Page 31: ...a en la pantalla El n mero que aparece en la posici n superior corresponde a la presi n sist lica y el n mero que aparece en la posici n inferior corresponde a la lectura de la presi n diast lica sist...

Page 32: ...to de su condici n Es por esto que es importante tomar mediciones diarias consistentes de su presi n sangu nea Las variaciones en las lecturas individuales s lo debe interpretarlas su m dico o un prof...

Page 33: ...esi n sangu nea en un lugar tranquilo Debe estar sentado y relajado 3 Evite fumar comer tomar medicamentos el consumo de alcohol o la actividad f sica 30 minutos antes de la lectura Si presenta se ale...

Page 34: ...sist lica 180 mmHg o Presi n arterial diast lica 110 mmHg Grado 2 de Hipertensi n Presi n arterial sist lica 160 179 mmHg o Presi n arterial diast lica 100 109 mmHg Grado 1 de Hipertensi n Presi n art...

Page 35: ...suarios controlar la presi n arterial con exactitud a n cu ndo los latidos del coraz n sean irregulares Cuando se detectan latidos del coraz n irregulares aparecer el cono IHB Fig 1 NOTA Solicite m s...

Page 36: ...ales cuando ya no se los utilice NOTA Cambie las bater as cuando aparezca el s mbolo de Poca Bater a en el visor o cuando el visor permanezca en blanco al presionar el bot n de ENCENDIDO 4 Se recomien...

Page 37: ...GADO POWER para marcar el ingreso 4 La fecha Fig 2 seguida de la hora y minutos titilar en forma intermi tente Repita los pasos 2 y 3 para establecer la configuraci n de fecha hora 5 Despu s de establ...

Page 38: ...a aur cula derecha del coraz n 4 Agrande o abra el brazalete extendiendo o enrollando la parte inferior del brazalete hacia la derecha Fig 2 Esto deber a abrir el brazalete creando un cilindro No exti...

Page 39: ...a tal como se muestra Fig 6 el monitor est listo El brazalete comenzar a inflarse autom ticamente 7 Una vez que el brazalete se haya inflado hasta el nivel de presi n adecuado comenzar la medici n 8 L...

Page 40: ...t n MEMORY para desplazarse por las mediciones almacenadas Fig 1 Una vez que se haya desplazado por todas las lecturas la unidad se apagar autom ticamente NOTA Cada banco de memoria puede guardar hast...

Page 41: ...ez que la flecha ha dejado de titilar la unidad est lista para tomar una medici n Aparece cuando trae una medici n de la memoria DIA SYS DIA SYS DIA SYS DESCRIPCI N DE LOS S MBOLOS DEL VISOR DESCRIPCI...

Page 42: ...as secciones C mo Colocar el Brazalete de Presi n Arterial y C mo Tomar una Lectura de Presi n Arterial Cambie las bater as Aparece el s mbolo EE Se movi o convers durante la medici n Refi rase a la s...

Page 43: ...nica Limpie el interior el lado que est en contacto con la piel del manguito con un pa o suave exprimido despu s de haberlo humedecido con alcohol et lico 75 90 y deje secar al aire 9 El usuario no d...

Page 44: ...V lvula de descarga de presi n autom tica Liberaci n r pida manual Memoria Visor con memoria incorporada que permite hasta 60 mediciones Apagado autom tico Alrededor de un minuto despu s de presionar...

Page 45: ...el 04 630 001 04 630 006 04 631 001 y 04 631 006 deben asegurarse de que se use en tal entorno Prueba de emisiones Cumplimiento Gu a sobre entornos electromagn tic Emisiones de RF CISPR 11 Grupo 1 04...

Page 46: ...re entornos electromagn ticos Descarga electrost tica electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Los pisos deben ser de madera cemento o azulejos cer micos S...

Page 47: ...1 y 04 631 006 deben asegurarse de que se use en tal entorno Prueba de INMUNIDAD Nivel de prueba seg n la norma IEC 60601 Nivel de cumplimiento Gu a sobre entornos electromagn ticos RF radiada IEC 610...

Page 48: ...de acuerdo con las indicaciones del fabricante del transmisor y d es la distancia de separaci n recomendada expresada en metros m Las resistencias de los campos de transmisores de RF fijos seg n lo d...

Page 49: ...ntensidad de campo obtenida en el lugar donde se utiliza el 04 630 001 04 630 006 04 631 001 y 04 631 006 excede el nivel de cumplimiento de RF correspondiente detallado m s arriba se debe verificar e...

Page 50: ...e 04 630 001 y 04 630 006 pueden ayudar a impedir la interferencia electromagn tica manteniendo una distancia m nima entre los equipos de comunicaci n por RF port tiles y m viles transmisores y 04 630...

Page 51: ...frecuencia del transmisor donde P corresponde a la potencia m xima de salida nominal del transmisor expresada en vatios V de acuerdo con las indicaciones del fabricante del transmisor NOTA 1 a 80 MHz...

Page 52: ...52 EJEMPLO DE REGISTRO DE PRESI N ARTERIAL...

Page 53: ...53 91 630 631 05 15 2015 HealthSmart International...

Page 54: ...54...

Reviews: