English
OVERVIEW
Thank you for choosing the P.T.Rail system from HealthCraft. Please read
and understand the instructions in this manual; keep manual for future
reference. It is your responsibility to see that your P.T.Rail system is
properly assembled, installed, and cared for. Failure to follow instructions in
this manual could result in serious injury or death. If you are not equipped
to undertake the outlined work, we would recommend that you have your
P.T.Rail system installed by a qualified contractor.
The P.T.Rail is a high quality, commercial grade, fold down rail for use
beside a toilet. The P.T.Rail features an offset handle to help reduce strain
when gripping the lower rail. The P.T.Rail is intended for moderate vertical
and horizontal load bearing to provide sitting and standing toilet support for
people with reduced mobility. The P.T.Rail is intended to bear weight ONLY
when the handle is in the down position; the handle is not intended to bear
weight in any other position. The P.T.Rail handle is intended to be folded up
when not in use. A friction hinge ensures that the handle will not
inadvertently lower when it is in the up position. The P.T.Rail is installed
with appropriate fasteners to a structural wall.
The Floor Mast is a floor mounted structural member intended to provide a
mounting surface for a P.T.Rail (in cases where wall fastening is not an
option). The Floor Mast is installed with appropriate fasteners to a
structural floor. The P.T.Rail is height adjustable on the Floor Mast. Must
use supplied fasteners to install P.T.Rail to Floor Mast.
The maximum user weight for this product is 400lbs / 181kgs. Note:
Product is not intended to support full body weight. The P.T.Rail and Floor
Mast are not to be used in any other way than described above.
WARRANTY
All HealthCraft Products are covered by a one year limited warranty with
the exception of hand grips which are subject to normal wear. Buyer
hereby indemnifies, agrees to hold harmless and defend HealthCraft
Products Inc. from and against any and all liabilities, claims, (founded and
unfounded), losses, damages, costs and expenses (including without
limitation consequential damages and reasonable professional fees)
resulting from buyers specification, application, or improper use of goods
described hereon; buyers omission or neglect.
HealthCraft Products Inc. does not assume any liability for damage
resulting from services performed by others or faulty installation, misuse or
misapplication of goods sold by HealthCraft Products Inc.
HealthCraft Products Inc. shall not be liable for prospective profits or
special, indirect, or consequential damages, or for the cost of any
corrective work done without HealthCraft Products Inc. prior approval.
HealthCraft Products Inc. total liability hereunder shall in no event exceed
the purchase price of the goods specified hereon.
Specifications subject to change without notice.
HealthCraft Products Inc.
2790 Fenton Road
Ottawa, Canada
K1T 3T7
DELIVERY CONTENTS - P.T.RAIL
FIGURE A.
1. P.T.Rail
2. Washer, 1/4", Lock (x5)
3. Washer, 1/4", Flat (x5)
4. 1/4" x 3" length Lag Bolt (x5)
DELIVERY CONTENTS - FLOOR MAST
FIGURE B.
5. Floor Mast
6. Clamp Plate
7. Bolt, 3/8"-16 x 3", S/S (x4)
8. Plastic Cover Sleeve (x4)
9. Vinyl Cap (x4)
P.T.RAIL INSTALLATION WARNINGS
FIGURE C.
1. P.T.Rail must be installed to walls that are structurally sound. Add wood
blocking if required.
2. If installation into concrete wall is required, we recommend sleeve or
wedge type anchors suitable for floor construction.
3. CAUTION - Do not install if structure behind wall is unknown.
4. CAUTION - Do not install into metal studs.
5. CAUTION - Only use rail for assistance in down position.
6. CAUTION - Pinch point between rail and wall plate.
7. CAUTION - Rail and bathtub are slippery when wet.
Supplied fasteners for wood structure only (minimum thread engagement
2" / 51mm). If using other fasteners, ensure screw retention force rating at
least 600lbs/272kgs.
FLOOR MAST INSTALLATION WARNINGS
FIGURE D.
1. Mast must be installed to floors that are structurally sound.
2. Ideal if mast can be aligned and secured directly to floor joist.
3. If desired installation location does not have adequate structure, add
wood blocking.
4. If installation into concrete floor is required, we recommend minimum
5/16" sleeve or wedge type anchor.
5. CAUTION - do not install if structure under floor is unknown.
SUGGESTED P.T.RAIL POSITIONS
FIGURE E.
To determine the optimal mounting location, we recommend holding the
rail in place and simulating the through the motions of a transfer (sitting,
standing, reaching ,etc.). Recommended rail height 32-34" / 813-864mm.
If one rail is installed on both sides of the toilet, suggested width between
top rails 28-32" / 711-813mm.
P.T.RAIL INSTALLATION INTO WOOD STRUCTURE
FIGURE F.
1. Hold P.T.Rail in desired location and ensure unobstructed rail travel in
all positions, and the space above rail is adequate to fold.
2. Using wall plate a template, mark off hole locations.
3. Drill pilot holes appropriate for fasteners.
4. Secure P.T.Rail to wall using appropriate fasteners.
FLOORMAST INSTALLATION INTO WOOD STRUCTURE FIGURE G.
1. Secure mast to floor using appropriate fasteners.
2. Attach the clamp plate to the back of the P.T.Rail.
3. Slide clamp plate with P.T.Rail over top of the mast.
4. With P.T.Rail at desired height, tighten bolts to secure in place.
5. Install vinyl caps to cover bolt heads.
TECHNICAL DATA
See Figure H.
PRODUCT USAGE
See Figure I.
EC
REP
Gate 88
Kanalgatan 45B
Skellefteå, Sweden
93123
10. Nut, 3/8"-16, S/S (x4)
11. Lock Washer, 3/8", S/S (x4)
12. Flat Washer, 3/8", S/S (x4)
13. Hex Key, 5/16"
10
Summary of Contents for P.T.Rail
Page 2: ...A 2 4 3 1 2 HW155 HW154 HW241...
Page 3: ...B 11 13 8 7 12 9 10 5 6 3 HW513 HW183 HW402 HW407 HW317 WR111C HW182...
Page 4: ...C Fe 4 5 6 7 1 2 3 4...
Page 5: ...E 32 34 813 864mm 28 32 711 813mm D 5 1 2 4 3 5 MIN 2 51mm...
Page 6: ...F 6 1 2 3 4 HW241 HW155 HW154...
Page 7: ...G 7 1 2 3 4 5 HW513 HW183 HW407 HW241 HW155 HW154 HW402 WR111C HW182...
Page 9: ...E I 1 3 3 1 3 3 9...