9
1600KL CALIBRATION
The calibration is performed using kilos (kg) or pounds (lb), according to the units of measure used
upon entering into the calibration path.
Operator Action
Display
1. Press ON/OFF button and immediately press and
hold HOLD/RELEASE button for 3-4 seconds
Calibration
Process
2. Press ENTER
Enter load weight
600.0
3. Using
keys and/or the keypad, set the
calibration load to at least 200 lbs (for best results
and greater accuracy, use a 600 lb weight). Press
ENTER
Zero calibration
Please clear the scale
4. Please clear the scale platform and press ENTER
Zero calibration
Please wait
5. Please wait without touching or shaking the scale
until the zero calibration process is finished
Weight calibration
Put: XXX.X
6. Load the scale platform with the required weight and
press ENTER
Weight calibration
Please wait
7. Please wait without touching or shaking the scale
until the calibration process is finished
Calibration
factor: X.XXXXX
8. Press ENTER
Calibration
Please clear the scale
9. Please remove weight from the scale platform
Rebooting
Please wait
10. Please wait until the scale resumes normal operation
Health o meter
ProPlus
MAINTENANCE
The following pages provide instructions for maintenance, cleaning, troubleshooting and operator
replaceable parts for the ProPlus
®
1600KL. Maintenance operations other than those described in
this manual should be performed by qualified service personnel.
MAINTENANCE
Before first use and after periods of non-use, check the scale for proper operation and function. If
the scale does not operate correctly, refer to qualified service personnel.
1.
Check overall appearance of the total scale for any obvious damage, wear and tear.
2.
Inspect AC adapter for cord cracking or fraying or for broken or bent prongs.
CLEANING
Proper care and cleaning is essential to ensure a long life of accurate and effective operation.
Disconnect the scale from the AC power source.
1.
Clean all external surfaces with a clean damp cloth or tissue. Mild antimicrobial or antibacterial
soap and water solution may be used. Dry with a clean soft cloth.
2.
Do not immerse the scale into cleaning or other liquid solution.
3.
Do not use Isopropyl Alcohol or other solutions to clean the display surface.
4. Do not use abrasive cleaners.