Health O Meter 553KG Instructions For Use Manual Download Page 6

 

POWERING THE SCALE / SET UP

 

Powering the Scale 

– Inserting Batteries 

1.  Locate the battery compartment underneath the scale base and remove battery 

compartment cover.  

2. 

Place 6 new AA batteries into the battery compartment. Make sure the ‘+’ and ‘-’ are 
positioned correctly. (Use only AA alkaline batteries). 

3.  Replace the battery compartment cover.  
Note: If the display shows “Lo”, replace the batteries as soon as possible.  

Warning: If the scale will not be used for some time, remove the batteries to avoid a 
safety hazard. Disposal of batteries must be performed in accordance with the prevailing 
national, regional or local regulations which apply to you

.

   

Note:

 

If “Lo” appears on the display, it indicates that the battery power is low. It does NOT 

indicate a problem with the weight measurement. 

 
Optional AC Adapter 

1.  Connect the AC adapter (not included) into the base of the scale and then into the outlet for 

power.  

 
 
 
 
 

2.  The LED next to the power jack will light up indicating there is power to the scale.  

Note:

 When the AC adapter is being used, the battery icon will not be displayed.  

 
 
Measuring Tape 

For your convenience a measuring tape is included with the scale. You may affix the tape to 
either the bottom or side of the weighing tray depending on your preference for visually 
determining the length of the baby, up to 24” or 58 cm.  

 
 
Powering Off the Scale / Auto-Off Feature 

To manually power off the scale, press the 

ON/OFF

 button. As an energy saving feature, the 

scale will automatically turn off after 2 minutes of inactivity. By default, this function is enabled. 
Follow the procedure below to deactivate or reactivate the auto off feature. 

 
Deactivate Auto-Off 

1.  While the scale is turned off, press and hold the 

RELEASE

 button. 

2.  While holding the 

RELEASE

 button, press the 

ON/OFF

 button to turn the scale on. 

3.  Keep holding down both buttons until  

“AOF=0” displays on the screen. AOF=0 confirms that 

Auto-Off has been deactivated.  

 
Activate Auto-Off 

1.  While the scale is turned off, press and hold the 

RELEASE

 button. 

2.  While holding the 

RELEASE

 button, press the 

ON/OFF

 button to turn the scale on. 

3.  Keep holding down both buttons 

until “AOF=1” displays on the screen. AOF=1 confirms that 

Auto-off is now reactivated. 

 
 
 

AC Adapter Jack 

Summary of Contents for 553KG

Page 1: ...Scale 44 lb 0 5 oz 20 kg x 0 01 kg User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM553KL_553KG Rev20180111_EngOnly 232791 IEC 60601...

Page 2: ...Connectivity Disposal 5 Powering the Scale Set Up 6 Operating Instructions 7 Maintenance 8 Calibration 9 Troubleshooting 10 Warranty 11 Note This scale has been factory calibrated and does not requir...

Page 3: ...Use of an unspecified adapter will void the warranty and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Pri...

Page 4: ...l Operating temperatures 50 F to 104 F 10 C to 40 C Storage Specifications 32 F to 122 F 0 C to 50 C Atmospheric Pressure 500hPa 1060hPa Humidity 10 90 RH Physical Dimensions Tray Size Length 13 330 m...

Page 5: ...nnectivity Solutions This connectivity helps minimize transcription errors when recording patient data For further information regarding how to connect this scale to other electronic devices please co...

Page 6: ...adapter is being used the battery icon will not be displayed Measuring Tape For your convenience a measuring tape is included with the scale You may affix the tape to either the bottom or side of the...

Page 7: ...e If Lo appears on the display it indicates that the battery power is low It does NOT indicate a problem with the weight measurement Zero Clear TARE Function The tare function allows you to place othe...

Page 8: ...be consumed Items such as clothing or towels place or dropped on the scale may cause it to activate To prevent run down of the batteries do not store or leave anything on the scale Cleaning Proper ca...

Page 9: ...can calibrate CAL followed by XXXXX Press the ZERO CLEAR button C 5 will flash on the display and then return to 0 C 5 then 0 Place 5 kg on the scale and when the calibration number stabilizes press t...

Page 10: ...Remove the patient from the weighing tray zero the scale and begin weighing process again 3 Scale is not placed on a level surface 3 Place the scale on a level surface and begin weighing process again...

Page 11: ...nadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get War...