CAUTIONS AND WARNINGS 

Intended use 

The Health o meter

®

 Professional 553KL and 553KG scales were intended to be used in a professional medical 

environment by trained medical staff. This product was designed to weigh infants that are laying or sitting on the 
tray. Do not modify the product or use it for anything other than its intended purpose. 
 

To prevent injury and damage to your scale, please follow these instructions very carefully. 

  Assemble and operate the scale per the enclosed user instructions. 

  For accurate weighing, this scale must be placed on a flat, stable surface. 

  For accurate weighing, verify before each use the proper operation according to the procedure 

described in this manual. 

  Do not transport the scale with a patient or object on the tray. 

  Do not exceed the weight capacity specified for the scale.  

  Do not use in the presence of flammable or explosive materials. 

  Ensure that the scale or AC adapter do not come in contact with liquids, excessive temperature, or 

excessive humidity. 

  If the scale becomes damaged, it should not be operated until properly serviced.  

 

Patient/caregiver safety 

  This scale is designed for static weighing of patients only. No scale should be used for patient 

transfer.  

  To prevent patient injury, the patient must be attended throughout the entire weighing event.   

 

When operating the scale with batteries: 

  I

f “Lo” appears on the display, replace the batteries or switch to the AC adapter as soon as possible.  

  When replacing / inserting batteries, be sure to use all new batteries. 

 
When operating the scale with the (optional) AC adapter: 

  Operate this scale exclusively with the AC adapter that is specified (ADPT30). Use of an unspecified 

adapter will void the warranty, and can pose a serious safety hazard.   

  Prior to using this scale, inspect the AC adapter cord for cracking / fraying, or for broken / bent plug 

prongs. 

  Prior to using this scale, ensure that the AC adapter is plugged into an outlet with the rated voltage 

appropriate for operation.   

  Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.  

  Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the 

scale.   

  Operating this scale at voltages and frequencies other than specified can damage the equipment 

and will void the warranty. 

 

This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 
93/42/EEC (EN 55011 Class A and EN 60601-1-2).  These limits are designed to provide reasonable 
protection against harmful interference in a typical medical installation. The equipment generates, uses 
and may radiate radio frequency energy and, if not installed and used in accordance with these 
instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that 
interference will not occur in a particular installation. If this equipment does cause harmful interference 
with other devices, which can be determined by turning the equipment off and on, the user is encouraged 
to try to eliminate the interference by one or more of the following measures: 
• 

Reorient or relocate the receiving device. 

• 

Increase the separation between the equipment. 

• 

Connect the equipment into an outlet on a circuit different from that to which the other device(s) is 

connected. 
• 

Consult Customer Service or a field service technician for help. 

 

In no event shall Pelstar, LLC be liable for damages or injuries arising out of or connected with the 
assembly, use, or misuse of this product 

Summary of Contents for 553KG

Page 1: ...Scale 44 lb 0 5 oz 20 kg x 0 01 kg User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2018 P N UM553KL_553KG Rev20180111_EngOnly 232791 IEC 60601...

Page 2: ...Connectivity Disposal 5 Powering the Scale Set Up 6 Operating Instructions 7 Maintenance 8 Calibration 9 Troubleshooting 10 Warranty 11 Note This scale has been factory calibrated and does not requir...

Page 3: ...Use of an unspecified adapter will void the warranty and can pose a serious safety hazard Prior to using this scale inspect the AC adapter cord for cracking fraying or for broken bent plug prongs Pri...

Page 4: ...l Operating temperatures 50 F to 104 F 10 C to 40 C Storage Specifications 32 F to 122 F 0 C to 50 C Atmospheric Pressure 500hPa 1060hPa Humidity 10 90 RH Physical Dimensions Tray Size Length 13 330 m...

Page 5: ...nnectivity Solutions This connectivity helps minimize transcription errors when recording patient data For further information regarding how to connect this scale to other electronic devices please co...

Page 6: ...adapter is being used the battery icon will not be displayed Measuring Tape For your convenience a measuring tape is included with the scale You may affix the tape to either the bottom or side of the...

Page 7: ...e If Lo appears on the display it indicates that the battery power is low It does NOT indicate a problem with the weight measurement Zero Clear TARE Function The tare function allows you to place othe...

Page 8: ...be consumed Items such as clothing or towels place or dropped on the scale may cause it to activate To prevent run down of the batteries do not store or leave anything on the scale Cleaning Proper ca...

Page 9: ...can calibrate CAL followed by XXXXX Press the ZERO CLEAR button C 5 will flash on the display and then return to 0 C 5 then 0 Place 5 kg on the scale and when the calibration number stabilizes press t...

Page 10: ...Remove the patient from the weighing tray zero the scale and begin weighing process again 3 Scale is not placed on a level surface 3 Place the scale on a level surface and begin weighing process again...

Page 11: ...nadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jurisdiction To get War...