Health O Meter 522KLWA User Instructions Download Page 3

 

 

CAUTIONS AND WARNINGS 

 

To prevent injury and damage to your scale, please follow these instructions very carefully: 

 

 

Do not transport the scale while platform is loaded.  Be sure the scale is tilted enough to assure it is rolling properly on its 
wheels when it is moved for cleaning or transported from one area to another. 

 

For accurate weighing the scale must be placed on a flat, stable surface. 

 

For accurate weighing, verify proper operation according to the procedure described in this manual before each use. 

 

Do not use in the presence of flammable materials. 

 

Operating at other voltages and frequencies than specified could damage the equipment. 

 

If the “LOW BAT” indicator activates, for accurate weighing replace the batteries or connect the scale to an AC power source 
as soon as possible. 

 

Only authorized, qualified personnel shall service the scale. 

 
This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 93/42/EEC (EN 
55011 Class A and EN 60601-1-2).  These limits are designed to provide reasonable protection against harmful interference in 
a typical medical installation. The equipment generates, uses and may radiate radio frequency energy and, if not installed and 
used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is 
no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with 
other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to eliminate the 
interference by one or more of the following measures: 

 

Reorient or relocate the receiving device.  

 

Increase the separation between the equipment. 

 

Connect the equipment into an outlet on a circuit different from that to which the other device(s) is connected. 

 

Consult Customer Service or a field service technician for help. 

 

DEFINITION OF SYMBOLS  

 

 

 

     

 

 

 

 

 

 

 

 

 

522KLWA SCALE SPECIFICATIONS  

 

Capacity and Resolution: 

0~9kg  d=5g   9~23kg  d=10g           0~20lb  d=0.2oz     20~50lb  d=0.5oz        

Power Requirements: 

6-AA cell 1.5V batteries  or 120-240V AC, 50-60Hz-9VDC Adapter Model ADPT30*  
BATTERIES AND ADAPTER ADPT30 INCLUDED 

Environmental: 

Operating Temperatures: 10°C to 40°C / 50°F to 104°F   
Atmospheric Pressure: 500hPa to 1060hPa 
Humidity: 10% to 95% RH 

Platform Dimensions: 

56.7 cm x 37.0 cm x 7.6 cm / 22” x 14 ” x 3” 

Scale Dimensions 

62.4 cm x 53.5 cm x 58.8 cm / 24 ½” x 21” x 23 1/8” 

 

 
STORAGE SPECIFICATIONS 

 

Storage Temperatures: 

0°C to 40°C / 32°F to 104°F 

Humidity: 

10% - 95% RH 

Atmospheric Pressure: 

500hPa – 1060hPa 

 

 
 

522KLWA SCALE CONNECTIVITY INFORMATION: 

 

This scale is designed to connect to other devices via a USB port, to allow acquisition of weight, from the 
scale, to help close a source of potential medical errors caused by incorrectly copying and then recording a 
weight taken.  This scale is EMR-EHR compatible.  For more information regarding how to connect this scale 
to your computer, a monitor or other device please contact Customer Service at one of the numbers listed on 
the next page.  

 

10C 

40C 

*WARNING:

  

 

For use with Adapter Model UE 15WCP-090050SPA/ADPT30 (IEC only).  
Complies with C22.2 No. 601.1-M90 and UL Std No. 60601-1. Type BF 
applied parts. Internally powered

  

 

Equipment is not suitable for use in the presence of flammable 
anesthetic mixture with air or with oxygen or nitrous oxide.

 

AC to DC power supply 
(9V DC, 500mA) 

 

DC Connector Polarity 

Temperature range 
(10°C to 40°C / 50°F to 104°F) 

Humidity range 
(10% - 95% RH) 

Atmospheric pressure range 
(500hPa – 1060hPa) 

Attention, Consult Accompanying Documents 

Type BF Applied Parts 

AC to DC power supply 

Summary of Contents for 522KLWA

Page 1: ...RICA DIGITAL BALANCE P DIATRIQUE NUM RIQUE 0 9 kg x 5g 9 23 kg 10g 0 20 lb 0 2 oz 20 50 lb x 0 5 oz Rev 20100722 232791 IEC 60601 1 C22 2 No 601 1 M90 UL 60601 1 PATENT D565446 US USER INSTRUCTIONS IN...

Page 2: ...rci d avoir achet cette balance Health o meter Professional Veuillez lire attentivement ce manuel et le garder port e de main pour pouvoir le consulter tout moment Health o meter Professional TABLE OF...

Page 3: ...Increase the separation between the equipment Connect the equipment into an outlet on a circuit different from that to which the other device s is connected Consult Customer Service or a field service...

Page 4: ...screw from the battery compartment cover and remove cover 3 Place 6 new AA alkaline batteries into the battery compartment Make sure the and are positioned correctly as shown on the bottom of the bat...

Page 5: ...ate and pull down the cable inside B and plug cable into the base C 2 Connect Upper Pillar B into Base C Secure Pillar to Base using the screws provided 3 Attach the Tray to the Base using the 2 screw...

Page 6: ...he scale on 3 START will briefly appear on the display followed by AOF 1 4 Auto off is now enabled De activating auto off 1 While the scale is turned off press and hold the HOLD RELEASE button 2 While...

Page 7: ...Plugged In 3 Electrical Outlet not functioning 1 Check Replace the batteries 2 Check Replace Power Supply Adapter 3 Use a different outlet LobAt shows in display The batteries are low Replace batterie...

Page 8: ...alternations Further the warranty does not cover Acts of God such as fire flood hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from...

Page 9: ...ipos Conecte el equipo a un tomacorriente en un circuito distinto al que est n conectados los otros instrumentos Consulte al Servicio de Atenci n al cliente o a un experto del servicio t cnico para so...

Page 10: ...pa del compartimento de las bater as y separe la tapa 3 Coloque 6 bater as alcalinas nuevas AA en el compartimiento Aseg rese de que las se ales y est n correctamente orientadas como se indica en el f...

Page 11: ...o en la base C 2 Conecte el pilar superior B a la base C Fije el pilar a la base con los tornillos provistos 3 Fije la bandeja a la base con los 2 tornillos provistos 4 Apriete los tornillos y c bralo...

Page 12: ...inicio seguido de AOF 1 4 El apagado autom tico ya est habilitado Desactivaci n del apagado autom tico 1 Mantenga presionado el bot n HOLD RELEASE retener liberar mientras la balanza est apagada 2 Mi...

Page 13: ...La balanza no se enciende 1 Problema de bater as 2 La fuente de alimentaci n no funciona o no est conectada 3 El tomacorriente no funciona 1 Revise o cambie las bater as 2 Revise o cambie el adaptado...

Page 14: ...as reparaciones o modificaciones no autorizadas Adem s la garant a tampoco cubre los da os causados por desastres naturales tales como incendios inundaciones huracanes y tornados La presente garant a...

Page 15: ...ugmenter l espace entre l quipement Relier l quipement une prise de courant d un autre circuit que celui sur lequel l autre les autres appareil s est sont reli s Consulter le service la client le ou u...

Page 16: ...mpartiment des piles et retirez le couvercle 3 Placez six piles alcalines AA neuves dans le compartiment des piles Veuillez vous assurer que le et le sont plac s correctement tel qu indiqu dans le fon...

Page 17: ...e c ble dans la base C 2 Connectez la colonne sup rieure B dans la base C Fixez la colonne la base l aide des vis fournies 3 Attachez le plateau la base l aide des 2 vis fournies 4 Serrez les vis et c...

Page 18: ...4 La fonction arr t automatique est maintenant activ e D sactivation de l arr t automatique 1 Pendant que la balance est teinte appuyez et tenez le bouton HOLD RELEASE maintenir rel cher 2 Tout en te...

Page 19: ...ion ou la balance n est pas branch e 3 La prise de courant ne fonctionne pas 1 V rifiez remplacez les piles 2 V rifiez remplacez l adaptateur de courant 3 Utilisez une autre prise LobAt piles faibles...

Page 20: ...cations non autoris es De plus la garantie ne couvre pas les catastrophes naturelles comme les incendies les inondations les ouragans et les tornades Cette garantie vous conf re des droits l gaux pr c...

Page 21: ...40V AC 50 60Hz 9VDC ADPT30 ADPT30 10 C 40 C 50 F 104 F 500hPa 1060hPa 10 95 56 7 x 37 0 x 7 6 22 x 14 x 3 62 4 x 53 5 x 58 8 24 x 21 x 23 1 8 0 C 40 C 32 F 104 F 10 95 500hPa 1060hPa 522KLWA USB EMR E...

Page 22: ...3 1 55 69 58 49 49 695 098 5132 39 026 968 2425 31 202 061 360 34 917 499 357 46 85 853 6551 44 207 365 6780 1 305 669 9003 49 7477 927186 27 11 777 7555 61 2 9638 3000 86 21 6327 9631 1 2 Phillips 3...

Page 23: ...23 B 4 1 B C 2 B C 3 2 4...

Page 24: ...LOCK 3 0 0 ON OFF HOLD RELEASE ZERO LB KG 2 1 HOLD RELEASE 2 HOLD RELEASE ON OFF 3 START AOF 1 4 1 HOLD RELEASE 2 HOLD RELEASE ON OFF 3 START AOF 0 4 KG LB KG LB 1 ON OFF 2 0 0 3 LB KG 4 LB KG 5 3 4...

Page 25: ...1 2 3 4 5 LB KG 5 10 15 20 6 1 2 3 1 2 3 LobAT 1 2 0 0 3 4 1 2 3 4 OVER 23 50 UNDER ZERO LB KG ZERO ON OFF CAL CAL XXXXX ZERO C 5 0 C 5 5 ZERO C 10 0 C 10 10 ZERO C 15 0 C 15 15 ZERO C 20 0 C 20 20 ZE...

Page 26: ...sional 2 Pelstar 22 ______ 10 100 000 PELSTAR LLC 11800 South Austin Avenue B Alsip IL 60803 USA www homscales com Health o meter Professional Pelstar LLC Welch Allyn Health o meter Sunbeam Products I...

Reviews: