Health O Meter 522KG User Instructions Download Page 7

P/N UM522KL_522KG Rev20190412                                     

 

ASSEMBLY INSTRUCTIONS (CONTINUED) 

 

1.  Carefully remove the contents 

of the carton and place each 
part on a flat, level, and dry 
surface. Remove the 4 column 
mounting screws and washers 
from the bottom of column and 
set aside for assembly step 5. 

 
 
 
 

2.  Remove the black adapter 

bracket from scale platform. 
Place the bracket on the 
bottom of the column with the 
adapter jack arrow sticker 
pointing towards the front of 
the column. 

 
 
 
 
 
 
 

3.  Place the scale platform on its 

side and place the column on 
its side, perpendicular to the 
platform. The front of the 
display head on the column 
should be facing forward and 
above the scale platform. 

 

4.  Locate the load cell connector 

cable inside the column and 
pull gently until the plug 
reaches the bottom of the 
column. Carefully pull the 
cable connector from the 
bottom of the column and plug 
into the RJ jack on the scale 
platform. 

Caution: Once 

connected, keep a firm grip 
on the columns to ensure 
the load cell connector cable 
is not pulled out from the RJ 
jack.

 

This cable connects the load 
cells to the display head and 
must be entirely plugged in 
for the scale to function 
properly.

 

 

 
 
 

Screws and washers 

RJ Jack 

Summary of Contents for 522KG

Page 1: ...G Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 522KL Health o meter Professional 522KG Pelstar LL...

Page 2: ...cations 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Operating Instructions 10 Maintenance and Cleaning 12 Calibration 13 Troubleshooting 14 Warranty 15 Note T...

Page 3: ...vely with the AC adapter provided by Health o meter Professional Use of an unspecified adapter will void the warranty and can pose a serious safety hazard For use with AC adapter model ADPT31 USA or C...

Page 4: ...ents Adapter Model No UE15WCP1 090050SPA Part No UE160714HKKK1 P Health o meter Professional Model ADPT30 Not Included Input 100 240V 50 60Hz 500mA Output 9 0V 0 5A Or 6 AA Batteries included Environm...

Page 5: ...638 3722 or visit www homscales com innovations connectivity Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or loca...

Page 6: ...as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid dam...

Page 7: ...s side and place the column on its side perpendicular to the platform The front of the display head on the column should be facing forward and above the scale platform 4 Locate the load cell connector...

Page 8: ...scale platform screw holes Using a Phillips head screwdriver tighten the screws to secure the column to the base of the platform 6 Locate the 2 M6 x 16mm screws included in the clear plastic bag Align...

Page 9: ...em is not corrected contact Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a pr...

Page 10: ...weight measuring unit follow the instructions below Locking the Weight Measuring Unit 522KL Only The weight measuring unit can be locked to only measure in KG or LB Follow the procedures below to lock...

Page 11: ...ASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the weighing tray After weight is removed from the tray the...

Page 12: ...Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild s...

Page 13: ...C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four numbers will appear When calibration numbers stabilize press the ZERO TARE button C 10 will appear on the...

Page 14: ...weight value Weighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not laying still Try to have the patient lay still The display sh...

Page 15: ...d hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jur...

Reviews: