TROUBLESHOOTING 

Troubleshooting 

Before contacting service personnel, refer to the following instructions to check and to 
correct any failures. 
 

Symptom 

Possible Cause 

Corrective Action 

Scale does not turn on 

1. Dead battery 
2. Faulty electrical outlet 
3. Bad power supply 

1. Replace batteries 
2. Use a different outlet 
3. Replace AC adapter 

Questionable weight or  
the scale does not zero 

1. External object 
interfering with the scale 

1. Remove interfering object 
from the scale 

 

2. The display did not 
show 

“0:0.0” before 

weighing 

2. Remove the patient from 
the weighing tray, zero the 
scale and begin weighing 
process again 

 

3. Scale is not placed on a 
level surface 

3. Place the scale on a level 
surface and begin weighing 
process again 

 

4. Scale is out of 
calibration 

4. Check weight with known 
weight value 

Weighing is performed 
but the weighing process 
takes too long and the 
weight does not lock on 
the display 

The patient is not laying 
still 

Try to have the patient lay 
still 

The display shows “Over” 
message 

The load on the scale 
exceeds the capacity 
 (50 lb / 23 kg) 

Remove the excess weight 
and use the scale according 
to its limits 

The display shows  
“Lo” message 

The batteries are depleted 

Replace batteries according 
to instructions 

Scale turns on but will not 
register a weight reading 

Load cell connector cable 
is not plugged in 

Refer to the instructions 
below to inspect the load cell 
cable.  
  

 
Inspecting the Load Cell Cable

 

 
 
 
 
 
 
 
 
 
 

Load Cell Cable  

Connected 

Load Cell Cable 

Not Connected 

1. Using a Phillips screwdriver, remove 
the door at the back of the platform 
base. 

2. Inspect the load cell connector cable to ensure 
it is completely plugged into the RJ jack. If cable 
was plugged in, unplug and plug in again until it 
clicks into place. After checking connection, 
replace door at the back of the platform base.  

Summary of Contents for 522KG

Page 1: ...G Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 522KL Health o meter Professional 522KG Pelstar LL...

Page 2: ...cations 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Operating Instructions 10 Maintenance and Cleaning 12 Calibration 13 Troubleshooting 14 Warranty 15 Note T...

Page 3: ...vely with the AC adapter provided by Health o meter Professional Use of an unspecified adapter will void the warranty and can pose a serious safety hazard For use with AC adapter model ADPT31 USA or C...

Page 4: ...ents Adapter Model No UE15WCP1 090050SPA Part No UE160714HKKK1 P Health o meter Professional Model ADPT30 Not Included Input 100 240V 50 60Hz 500mA Output 9 0V 0 5A Or 6 AA Batteries included Environm...

Page 5: ...638 3722 or visit www homscales com innovations connectivity Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or loca...

Page 6: ...as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid dam...

Page 7: ...s side and place the column on its side perpendicular to the platform The front of the display head on the column should be facing forward and above the scale platform 4 Locate the load cell connector...

Page 8: ...scale platform screw holes Using a Phillips head screwdriver tighten the screws to secure the column to the base of the platform 6 Locate the 2 M6 x 16mm screws included in the clear plastic bag Align...

Page 9: ...em is not corrected contact Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a pr...

Page 10: ...weight measuring unit follow the instructions below Locking the Weight Measuring Unit 522KL Only The weight measuring unit can be locked to only measure in KG or LB Follow the procedures below to lock...

Page 11: ...ASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the weighing tray After weight is removed from the tray the...

Page 12: ...Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild s...

Page 13: ...C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four numbers will appear When calibration numbers stabilize press the ZERO TARE button C 10 will appear on the...

Page 14: ...weight value Weighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not laying still Try to have the patient lay still The display sh...

Page 15: ...d hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jur...

Reviews: