Health O Meter 522KG User Instructions Download Page 3

P/N UM522KL_522KG Rev20190412                                     

 

CAUTIONS AND WARNINGS 

Intended Use 

The Health o meter

®

 Professional 522KL/522KG scale is intended to be used in a professional medical 

environment by trained medical staff. This product was designed to weigh infants that are laying or sitting on the 
tray. The intended use of the included measuring tape is to measure patient length. Do not modify the product 
or use it for anything other than its intended purpose. 
 

To prevent injury and damage to your scale, please follow these instructions very carefully. 

  Assemble and operate the scale per the enclosed user instructions. 

  For accurate weighing, this scale must be placed on a flat, stable surface. 

  For accurate weighing, verify before each use the proper operation according to the procedure 

described in this manual. 

  Do not transport the scale with a patient or object on the tray. 

  Do not exceed the weight capacity specified for this scale.  

  Do not use in the presence of flammable or explosive materials. 

  Ensure that the scale or AC adapter do not come in contact with liquids, excessive temperature, or 

excessive humidity. 

  If the scale becomes damaged, it should not be operated until properly serviced.  

 

Patient/caregiver safety 

  This scale is designed for static weighing of patients only. No scale should be used for patient 

transfer.  

  To prevent patient injury, the patient must be attended throughout the entire weighing event.  

  Patient should wear lightweight covering or a paper tray liner should be used during weighing event.    

 

When operating the scale with batteries: 

  I

f the “LO” indicator activates, replace the batteries or switch to the AC adapter as soon as possible.  

  When replacing / inserting batteries, be sure to use all new batteries. 

 
When operating the scale with the (optional) AC adapter: 

  Operate this scale exclusively with the AC adapter provided by Health o meter Professional.  Use of 

an unspecified adapter will void the warranty, and can pose a serious safety hazard.   

  For use with AC adapter model ADPT31 (USA or Canada) or ADPT30 (International) (not included)

 

  Prior to using this scale, inspect the AC adapter cord for cracking / fraying, or for broken / bent plug 

prongs. 

  Prior to using this scale, ensure that the AC adapter is plugged into an outlet with the rated voltage 

appropriate for operation.   

  Ensure that the AC adapter outlet is wired to a circuit breaker or other protected power source.  

  Unplug the AC adapter, and carefully store both the adapter and adapter cord before moving the 

scale.   

  Operating this scale at voltages and frequencies other than specified can damage the equipment 

and will void the warranty. 

 

This equipment has been tested and found to comply with the EMC limits for the Medical Device Directive 
93/42/EEC (EN 55011 Class A and EN 60601-1-2).  These limits are designed to provide reasonable 
protection against harmful interference in a typical medical installation. The equipment generates, uses 
and may radiate radio frequency energy and, if not installed and used in accordance with these 
instructions, may cause harmful interference to other devices in the vicinity. There is no guarantee that 
interference will not occur in a particular installation. If this equipment does cause harmful interference 
with other devices, which can be determined by turning the equipment off and on, the user is encouraged 
to try to eliminate the interference by one or more of the following measures: 
• 

Reorient or relocate the receiving device. 

• 

Increase the separation between the equipment. 

• 

Connect the equipment into an outlet on a circuit different from that to which the other device(s) is 

connected. 
• 

Consult Customer Service or a field service technician for help. 

 

In no event shall Pelstar, LLC be liable for damages or injuries arising out of or connected with the 

assembly, use, or misuse of this product. 

Summary of Contents for 522KG

Page 1: ...G Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 522KL Health o meter Professional 522KG Pelstar LL...

Page 2: ...cations 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Operating Instructions 10 Maintenance and Cleaning 12 Calibration 13 Troubleshooting 14 Warranty 15 Note T...

Page 3: ...vely with the AC adapter provided by Health o meter Professional Use of an unspecified adapter will void the warranty and can pose a serious safety hazard For use with AC adapter model ADPT31 USA or C...

Page 4: ...ents Adapter Model No UE15WCP1 090050SPA Part No UE160714HKKK1 P Health o meter Professional Model ADPT30 Not Included Input 100 240V 50 60Hz 500mA Output 9 0V 0 5A Or 6 AA Batteries included Environm...

Page 5: ...638 3722 or visit www homscales com innovations connectivity Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or loca...

Page 6: ...as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid dam...

Page 7: ...s side and place the column on its side perpendicular to the platform The front of the display head on the column should be facing forward and above the scale platform 4 Locate the load cell connector...

Page 8: ...scale platform screw holes Using a Phillips head screwdriver tighten the screws to secure the column to the base of the platform 6 Locate the 2 M6 x 16mm screws included in the clear plastic bag Align...

Page 9: ...em is not corrected contact Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a pr...

Page 10: ...weight measuring unit follow the instructions below Locking the Weight Measuring Unit 522KL Only The weight measuring unit can be locked to only measure in KG or LB Follow the procedures below to lock...

Page 11: ...ASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the weighing tray After weight is removed from the tray the...

Page 12: ...Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild s...

Page 13: ...C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four numbers will appear When calibration numbers stabilize press the ZERO TARE button C 10 will appear on the...

Page 14: ...weight value Weighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not laying still Try to have the patient lay still The display sh...

Page 15: ...d hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jur...

Reviews: