MAINTENANCE AND CLEANING 

Maintenance 

The following pages provide instructions for maintenance, cleaning, calibrating, and 
troubleshooting of your scale. Maintenance operations other than those described in this 
manual should be performed by qualified service personnel. 

Caution: Before first use, or after long periods of non-use, check the scale for proper 
operation and function. If the scale does not operate correctly, refer to qualified service 
personnel. 

1. Check overall appearance of the total scale for any obvious damage, wear, and tear. 
2. Inspect the AC adapter cord for cracking or fraying, or for broken/bent prongs. 
 

Note:

 This scale is an extremely sensitive weighing device. If the auto-off function is disabled, 

battery power will be consumed. 
 

Cleaning and Disinfecting 

Proper care and cleaning is essential to ensure a long life of accurate and effective weighing. 

 
Caution: Disconnect the scale from the AC adapter power source prior to cleaning the 
unit.          
 

1.  Health o meter

®

 Professional recommends using one of the following solutions on a soft 

cloth or disposable wipe: 
●  mild soap and water solution 
●  70% isopropyl alcohol 
●  solution with 1-5% hydrogen peroxide concentration 
After cleaning/disinfecting, wipe with a cloth dampened with water and then with a clean dry 
cloth. To avoid residue build-up or damp surfaces, ensure the screen and scale parts are 
completely dry after cleaning. 

2. 

Never use rough or abrasive materials to clean the scale, as these will damage the scale’s 
finish.  

3.  Do not submerge the scale in water or any other liquid.  

4.  Do not pour or spray fluids directly on the scale. 
  

Replacement Parts 

Replacement parts may be available if a piece of the scale needs to be replaced. Contact Health o 
meter Professional Scales Customer Service at 1-800-815-6615 to inquire about the availability of 
these replacement parts. 

Part # 

Description 

522KL/524KLKEYPAD 

522KL Keypad 

522TRAY 

Tray for 522KL/522KG scale 

ADPT31 

Power adapter (USA and Canada) 

ADPT30 

Power adapter (International) 

Summary of Contents for 522KG

Page 1: ...G Digital Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55th St McCook IL 60525 7110 USA www homscales com Health o meter Professional 522KL Health o meter Professional 522KG Pelstar LL...

Page 2: ...cations 4 Certifications Connectivity Disposal 5 Assembly Instructions 6 Powering the Scale 9 Operating Instructions 10 Maintenance and Cleaning 12 Calibration 13 Troubleshooting 14 Warranty 15 Note T...

Page 3: ...vely with the AC adapter provided by Health o meter Professional Use of an unspecified adapter will void the warranty and can pose a serious safety hazard For use with AC adapter model ADPT31 USA or C...

Page 4: ...ents Adapter Model No UE15WCP1 090050SPA Part No UE160714HKKK1 P Health o meter Professional Model ADPT30 Not Included Input 100 240V 50 60Hz 500mA Output 9 0V 0 5A Or 6 AA Batteries included Environm...

Page 5: ...638 3722 or visit www homscales com innovations connectivity Scale Disposal This Health o meter Professional scale must be disposed of properly as electronic waste Follow the national regional or loca...

Page 6: ...as you unpack the parts for assembly To prevent scratching any components carefully remove each assembly from the carton and unwrap the packing materials Set the carton aside for storage To avoid dam...

Page 7: ...s side and place the column on its side perpendicular to the platform The front of the display head on the column should be facing forward and above the scale platform 4 Locate the load cell connector...

Page 8: ...scale platform screw holes Using a Phillips head screwdriver tighten the screws to secure the column to the base of the platform 6 Locate the 2 M6 x 16mm screws included in the clear plastic bag Align...

Page 9: ...em is not corrected contact Health o meter Professional Scales Customer Service at 1 800 815 6615 Note If Lo appears on the display it indicates that the battery power is low It does NOT indicate a pr...

Page 10: ...weight measuring unit follow the instructions below Locking the Weight Measuring Unit 522KL Only The weight measuring unit can be locked to only measure in KG or LB Follow the procedures below to lock...

Page 11: ...ASE button To release press the HOLD RELEASE button again 5 After the weight measurement has been taken and is recorded remove the baby from the weighing tray After weight is removed from the tray the...

Page 12: ...Disconnect the scale from the AC adapter power source prior to cleaning the unit 1 Health o meter Professional recommends using one of the following solutions on a soft cloth or disposable wipe mild s...

Page 13: ...C 5 will flash on the display followed by 0 or 1 C 5 then 0 or 1 Place 5 kg on the scale Four numbers will appear When calibration numbers stabilize press the ZERO TARE button C 10 will appear on the...

Page 14: ...weight value Weighing is performed but the weighing process takes too long and the weight does not lock on the display The patient is not laying still Try to have the patient lay still The display sh...

Page 15: ...d hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to province or jurisdiction to jur...