TROUBLESHOOTING
Troubleshooting
Before contacting service personnel, refer to the following instructions to check and to correct
any failures.
Symptom Possible
Cause
Corrective
Action
Scale does not turn on
1. Dead battery
2. Faulty electrical outlet
3. Bad power supply
1. Replace batteries
2. Use a different outlet
3. Replace AC adapter
Questionable weight or
the scale does not zero
1. External object interfering
with the scale
1. Remove interfering object
from the scale
2. The display did not show
“0.0” before weighing
2. Ask the patient to get off the
scale, zero the scale and begin
weighing process again
3. Scale is not placed on a
level floor
3. Place the scale on a level
floor and begin weighing
process again
4. Scale is out of calibration
4. Check weight with known
weight value
5. Improper Tare
5. Place the item to be tared on
the scale. Press
REWEIGH
.
Once the weight of the item is
displayed, press
TARE
. Place
the patient and the tared item
back on the scale. Press
REWEIGH
again
Weighing is performed but
the weighing process takes
too long and the weight
does not lock on the
displayed
The patient is not sitting or
standing still
Ask the patient to stand or sit
still
The display shows “Over”
message
The load on the scale
exceeds the capacity
(800 lb / 363 kg)
Remove the excess weight and
use the scale according to its
limits
The display shows
“Lo bAtt” message
The batteries are empty
Replace batteries according to
instructions
Summary of Contents for 2400KL
Page 1: ...1 Model 2400KL Digital Portable Wheelchair Scale User Instructions P N UM2400KL Rev 1 032212 ...
Page 9: ...9 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 1 Figure 2 ...
Page 10: ...10 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 3 Figure 4 Load cell port ...
Page 12: ...12 2400KL ASSEMBLY INSTRUCTIONS CONTINUED Figure 7 ...
Page 16: ...16 SETUP BATTERIES TRANSPORT CONTINUED Figure 10 Figure 11 ...