SET UP INSTRUCTIONS
Activating/Deactivating Sound
A tone sounds whenever a weight is locked
or a button is pressed. (Factory setting is “on”) The scale
can be set to stay silent or beep while operating the scale. Follow this procedure to set the preferred
option.
1. Press
ON/ZERO
and wait for the scale to display “0.0”.
2. Hold down
MENU
and press
RECALL
. “Set-0” will be displayed and the “0” will be flashing.
3. Press the
button repeatedly until “Set-5” appears. Press
ENTER
.
Note: the button must be pressed quickly after
“Set-0” appears or the scale will return to zero.
4.
“11011” will appear on the display. The sound setting is changed with the first digit “11011”.
5.
With the first digit flashing on the display, select “1” or “0” for your preferred option. Use the
and
buttons to input your desired setting.
a.
To keep sounds on, select “1”. Example: Sound on = 11011
b.
To turn sound off, select “0”. Example: Sound off = 01011
6. Press
ENTER
when your selection is complete. Press
ENTER
to advance to the next
digit / setting. If no other settings need to be changed press
ENTER
repeatedly until the scale
returns to zero.
Programming Auto-Off Time
The 2210KL / 2210KL4 has been pre-set at the factory to shut down after one minute of inactivity.
This can be changed to any increment (from 12 seconds to 25 minutes) to suit your needs.
1. Press
ON/ZERO
and wa
it for the scale to display “0.0”.
2. Hold down
MENU
and press
RECALL
. “Set-0” will be displayed and the “0” will be flashing.
3. Press the button repeatedly until
“Set-3” appears. Press
ENTER
.
Note: the button must be pressed quickly after
“Set-0” appears or the scale will return to zero.
4. The Auto-Off time currently programmed in the scale will display as minutes and tenths of minutes
(i.e.
“03.2”); the far left digit will be flashing. Use the and buttons to set the desired value,
press
ENTER
to advance to the next digit. For example, setting the time to
“00.5” will turn the
scale off after 30 seconds of no use;
“02.0” will turn the scale off after 2 minutes; “25.0” will turn
the scale off after 25 minutes.
5. Press
ENTER
when complete to save and return to normal weighing operation.
Unit of Measure Configuration (LB / GM or LB / KG)
The unit of measure may be displayed in lb / grams (i.e. 2541 gram) or lb / kilograms (2.541 kg).
(Factory setting is pounds / grams) Follow this procedure to set the preferred option.
1. Press
ON/ZERO
and wait for the scale to display “0.0”.
2. Hold down
MENU
and press
RECALL
. “Set-0” will be displayed and the “0” will be flashing.
3.
Press the button repeatedly until “Set-5” appears. Press
ENTER
.
Note: the button must be pressed quickly after
“Set-0” appears or the scale will return to zero.
4.
“11011” will appear on the display. The unit of measure setting is changed with the fifth digit
“11011”.
5. Press
ENTER
until the fifth digit is flashing. Use the
and
buttons to input your desired
setting.
a.
To weigh in pounds/kilograms, select “0”.
b. To weigh in pounds / grams select
“1”.
Example: LB / KG display = 11010
LB / GM display = 11011
6. Press
ENTER
when your selection is complete and the scale will return to zero.