CALIBRATION 

Your scale has been factory calibrated and does not require calibration prior to use. To check for 
accuracy, place an accurately known certified weight (3,000 to 5,000 grams recommended) on the 
platform. If the display reads other than the known weight value by +/- 5 grams, Health o meter 
Professional recommends calibrating the scale.  
 
Calibration must be done in grams or kilograms using certified weights.  

 

IMPORTANT NOTE

:  A certified weight is a weight that has been accurately calibrated and its 

license is traceable to NBS weights & measures.  Weight room free weights are 

not 

suitable to use 

as check weights as they may vary from their stated value.   

 

The 2210KL/2210KL4 is an ultra-accurate scientific instrument.  Environmental conditions during the 
calibration process can cause slight deviations in the calibration results.  While these deviations will 
not affect the overall performance of the scale, we recommend the following procedure for the most 
accurate calibration results: 

 

1.  Press 

ON/ZERO

 

and wait for the display to show “0.0”.  

 

2.  With no weight on the platform, begin the calibration procedure. 

NOTE:

 Ensure the scale is on 

for a minimum of 2 minutes before beginning calibration. If the Auto Off time is programmed for 
less than 2 minutes, either occasionally tap the scale so it does not power off or change the Auto 
Off time to longer than 2 minutes.  

 

3.  The calibration security code must be entered before attempting to calibrate the scale. Hold 

down 

MENU

 and press 

RECALL

. Release both buttons. “Set-0” will be displayed and the “0” will 

be flashing. 

 

4.  Press the       

button repeatedly until “Set-2” appears. Press 

ENTER

. Note: the       button must 

be pressed qui

ckly after “Set-0” appears or the scale will return to zero. 

 

5.  Use the     and 

 

keys to input “125” as the calibration security code. Press 

ENTER

 to advance 

to the next digit. After entering the third digit, press 

ENTER

 and the scale will return to normal 

weighing mode. The security code needs to be entered only once, unless the scale is powered 
off. Access to calibration is denied if the proper code is not entered. 

 

6. 

After “0” appears on the display, wait 5 seconds then press 

ON/ZERO

 to zero the scale. 

 

7.  Place the certified weight onto the scale and wait for the weight measurement to lock. Press 

REWEIGH t

hen quickly hold down 

MENU

 and press 

RECALL

. Release both buttons. 

Note:

 this 

button sequence must be done quickly to ensure the weight does not lock after pressing 

REWEIGH

.   

 

8. 

“Set-0” will be displayed and the “0” will be flashing.Press the      button repeatedly until “Set-6” 
appears. Press 

ENTER

 

9.  A weight value will display representing the current calculation of the known weight. The grams 

digit will be flashing. Use the       and       keys to input the proper value of the known weight. 
Press 

ENTER

 to advance to the next digit. After entering the weight value, press 

ENTER.            

“-CAL-“ will appear on the display followed by dashes while the scale calibrates. 

 

10. The calibration weight value will appear on the display. Remove weight, wait 5 seconds, then 

press 

ON/ZERO

. Wait for “0.0” to display and verify calibration by placing the same known 

weight back on the scale. Calibration is now complete. For additional questions or technical 
support contact Customer Service at 1-800-815-6615. 

 
Calibration Error Code

 

“Err-4” 

Scale is below minimum calibration weight.  Use at least 3000 grams to 
calibrate the scale. 

Summary of Contents for 2210KL

Page 1: ...4 Rev20190312_Web 1 Model 2210KL 2210KL4 High Resolution Digital Neonatal Pediatric Tray Scale User Instructions PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pelstar LLC 2...

Page 2: ...s Scale 4 Certifications Disposal 5 Assembly Instructions 6 Set Up Instructions 8 Operating Instructions 11 Calibration 14 Troubleshooting 15 Maintenance 16 Warranty 17 Note This scale has been factor...

Page 3: ...se in the presence of flammable or explosive materials If the scale becomes damaged it should not be operated until properly serviced Do not transport the scale with a patient or object on the scale E...

Page 4: ...apacity and Resolution 20 000 grams 20 kg 45 lb Resolution 1 g to 10 000 g 5 g 10 001 to 20 000 g 0 1 oz 0 45 lb Power Requirements 6 C Cell Alkaline batteries Environmental Operating temperatures 68...

Page 5: ...ou for disposal of electronic waste or batteries Do not dispose of this device in the domestic waste stream Pelstar LLC has been officially certified as the manufacturer of Health o meter Professional...

Page 6: ...ssible after receipt of the package The following information details what you will find inside the main carton as you unpack the parts for assembly To prevent scratching any components carefully remo...

Page 7: ...ing the 6 screws Once cover is centered tighten screws fully Ensure the scale cover is replaced in the same orientation that it was removed The cover only fits properly in one direction or the tray wi...

Page 8: ...L Set 0 will be displayed and the 0 will be flashing 3 Press the button repeatedly until Set 3 appears Press ENTER Note the button must be pressed quickly after Set 0 appears or the scale will return...

Page 9: ...ing the lock procedure 1 Press ON ZERO and wait for the scale to display 0 0 2 Hold down MENU and press RECALL Set 0 will be displayed and the 0 will be flashing 3 Press the button repeatedly until Se...

Page 10: ...sed quickly after Set 0 appears or the scale will return to zero 4 10010 or 10011 will appear on the display The lock setting is changed with the second digit 10011 6 Press ENTER until the second digi...

Page 11: ...O Press to turn scale on or reset to zero TARE Used to perform the tare function and deducts the weight of an object on the tray LB GM Toggles the unit of measurement between pounds and grams in LB GM...

Page 12: ...tomatically performs this subtraction avoiding the need for manual calculations 1 Turn the scale on and wait for the display to read 0 2 Place the object i e blanket on the tray The weight of the obje...

Page 13: ...ection uses a proprietary technology that automatically provides accurate weight measurement when there is motion on the tray If there was considerable motion during the weight measurement motion dete...

Page 14: ...ll be flashing 4 Press the button repeatedly until Set 2 appears Press ENTER Note the button must be pressed quickly after Set 0 appears or the scale will return to zero 5 Use the and keys to input 12...

Page 15: ...es and check their positioning 2 Check that all new and fresh batteries are properly inserted into the battery tray Incorrect weight readings 1 Tray is touching a wall or other object 2 Tray is not le...

Page 16: ...oft cloth or disposable wipe mild soap and water solution 70 isopropyl alcohol solution with 1 5 hydrogen peroxide concentration After cleaning disinfecting wipe with a cloth dampened with water and t...

Page 17: ...isasters such as fire flood hurricanes and tornadoes This warranty gives you specific legal rights and you may also have other rights that vary from country to country state to state province to provi...

Page 18: ......

Page 19: ...312_Web 1 Modelo 2210KL 2210KL4 B scula Digital de Bandeja de Alta Resoluci n Neonatal Pedi trica Instrucciones del Usuario PELSTAR LLC 9500 West 55 th St McCook IL 60525 7110 USA www homscales com Pe...

Page 20: ...la 4 Certificaciones Disposici n 5 Instrucciones de Ensamblaje 6 Instrucciones de Instalaci n 8 Instrucciones de Funcionamiento 11 Calibraci n 14 Soluci n de Problemas 15 Mantenimiento 16 Garant a 17...

Page 21: ...en presencia de materiales inflamables o explosivos Si la b scula se da a no debe utilizarse hasta que se le d el servicio adecuado No transporte la b scula con un paciente u objeto sobre la misma Ase...

Page 22: ...ci n 20 000 gramos 20 kg 45 lb Resoluci n 1 g hasta 10 000 g 5 g de 10 001 a 20 000 g 0 1 oz 0 45 lb Requerimientos de Energ a 6 bater as Alcalinas de Celda C Ambiente Temperaturas de funcionamiento 6...

Page 23: ...d para la elminaci n de desechos electr nicos o bater as No se deshaga de este dispositivo como desecho dom stico Pelstar LLC se ha certificado oficialmente como el fabricante de los dispositivos m di...

Page 24: ...n siguiente detalla lo que usted encontrar dentro del empaque de cart n principal al desempacar los componentes para su ensamblado Para evitar rasgar cualquiera de los componentes extraiga cuidadosame...

Page 25: ...apriete completamente los tornillos Aseg rese de que la cubierta de la b scula se reemplace en la misma orientaci n tal como fue retirada La cubierta solamente entra en una direcci n o la bandeja no a...

Page 26: ...o el bot n MEN y presione el bot n RECALL Se visualizar Set 0 y destellar 0 3 Presione el bot n repetidamente hasta que aparezca Set 3 presione ENTER Nota el bot n debe ser presionado inmediatamente l...

Page 27: ...cedimiento de bloqueo 1 Presione el bot n ON ZERO y espere a que la b scula muestre 0 0 2 Mantenga presionado MENU y presione RECALL Se visualizar Set 0 y destellar 0 3 Presione el bot n repetidamente...

Page 28: ...iatamente luego de que aparezca Set 0 o la b scula regresar a cero 4 10010 o 10011 aparecer en la pantalla El ajuste de bloqueo se cambia con el segundo d gito 10011 5 Presione ENTER hasta que destell...

Page 29: ...r la b scula o reajustar a cero TARE TARA Se usa para realizar la funci n de tara y resta el peso de un objeto sobre la bandeja LB GM Cambia la unidad de medici n entre libras y gramos en el modoLB GM...

Page 30: ...itando la necesidad de c lculos manuales 1 Encienda la b scula y espere a que aparezca 0 en la pantalla 2 Coloque el objeto por ejemplo la manta sobre la bandeja Se mostrar el peso del objeto 3 Presio...

Page 31: ...Detecci n de Movimiento La detecci n del movimiento utiliza una tecnolog a propia que proporciona autom ticamente la medida exacta del peso cuando hay movimiento en la bandeja Si hubo movimiento consi...

Page 32: ...ente hasta que aparezca Set 2 Presione ENTER Nota el bot n debe ser presionado inmediatamente luego de que aparezca Set 0 o la b scula regresar a cero 5 Use las teclas y para ingresar 125 como el c di...

Page 33: ...comprobar su posici n 4 Compruebe que las bater as sean recientes y nuevas y que est n insertadas apropiadamente en la bandeja de bater as Lecturas de peso incorrectas 3 La bandeja est tocando la pare...

Page 34: ...oluci n suave de jab n y agua alcohol isoprop lico al 70 soluci n de per xido de hidr geno concentraci n 1 5 Despu s de la limpieza desinfecci n pase un pa o humedecido con agua y luego un pa o limpio...

Page 35: ...so incorrecto del producto uso de voltaje o corriente incorrectos uso contrario a las instrucciones de operaci n abuso incluyendo manipulaci n da o en el transporte o la reparaci n o alteraciones no a...

Reviews: