Load‐carrying capacity
44
A03452_15_X00_00
HBM: public
T40B
9
Load‐carrying capacity
Nominal (rated) torque can be exceeded statically up to the limit torque. If the
nominal (rated) torque is exceeded, additional irregular loading is not permissi
ble. This includes longitudinal forces, lateral forces and bending moments.
Limit values can be found in
Chapter 13 “Specifications”, page 50
Measuring dynamic torque
The torque flange can be used to measure static and dynamic torque. The fol
lowing apply to the measurement of dynamic torque:
S
The T40B calibration performed for static measurements is also valid for
dynamic torque measurements.
S
The natural frequency f
0
of the mechanical measuring arrangement
depends on the moments of inertia J
1
and J
2
of the connected rotating
masses and the torsional stiffness of the T40B.
Use the equation below to approximately determine the natural frequency f
0
of
the mechanical measuring arrangement:
f
0
+
1
2
p
·
c
T
·
ǒ
1
J
1
)
1
J
2
Ǔ
Ǹ
f
0
= natural frequency in Hz
J
1,
J
2
= mass moment of inertia in kgVm
2
c
T
= torsional stiffness in NVm/rad
S
The permissible mechanical oscillation width (peak‐to‐peak) can also be
found in the specifications.
Summary of Contents for T40B
Page 1: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Page 3: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Page 70: ...Supplementary technical information 68 A03452_15_X00_00 HBM public T40B ...
Page 71: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Page 163: ...Dimensions Abmessungen T40B A03452_15_X00_00 HBM public 93 ...