Electrical connection
T40B
A03452_15_X00_00
HBM: public
31
6
Electrical connection
6.1
General information
S
With extension cables, make sure that there is a proper connection with
minimum contact resistance and good insulation.
S
All cable connectors or swivel nuts must be fully tightened.
Important
Transducer connection cables from HBM with plugs attached are identified in
accordance with their intended purpose (Md or n). When cables are shortened,
inserted into cable ducts or installed in control cabinets, this identification can
get lost or become concealed. So the cables must be marked beforehand, just
in case.
6.2
EMC protection
Important
The transducers are EMC‐tested in accordance with EC directives and identi
fied by CE certification. However, you must connect the shield of the connec
tion cable on the shielding electronics enclosure in order to achieve EMC pro
tection for the measuring chain.
Special electronic coding methods are used to protect the purely digital signal
transmission between the transmitter head and the rotor from electromagnetic
interference.
The cable shield is connected with the transducer housing. This encloses the
measurement system (without the rotor) in a Faraday cage when the shield is
laid flat at both ends of the cable. With other connection techniques, an
EMCproof shield should be applied in the wire area, and this shielding should
also be connected extensively (also see HBM Greenline Information, brochure
i1577).
Summary of Contents for T40B
Page 1: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Page 3: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Page 70: ...Supplementary technical information 68 A03452_15_X00_00 HBM public T40B ...
Page 71: ...Mounting Instructions Montageanleitung English Deutsch T40B ...
Page 163: ...Dimensions Abmessungen T40B A03452_15_X00_00 HBM public 93 ...