6.2 Configuration
The configuration defines which data content is exchanged in the cyclical data
communication. The following data is available for selection:
Input values:
Name
Meaning
Length
Gross
gross measured value
1 or 2
words
Net
net measured value (gross minus tare value)
1 or 2
words
Max
contents of the maximum store
1 or 2
words
Min
contents of the minimum store
1 or 2
words
Pk
−
Pk
peak
−
to
−
peak, difference between max and min
1 or 2
words
Status1
status word with status of the limit switches and gen. error
bits
1 word
Status2
double status word with differentiated error flagging
2 words
Output values:
Name
Meaning
Length
Control word
control word for triggering taring, zeroing, clearing the peak
value store, parameter set selection, etc.
1 word
Limit1
level at which limit switch 1 responds
1 or 2
words
Limit2
level at which limit switch 2 responds
1 or 2
words
Limit3
level at which limit switch 3 responds
1 or 2
words
Limit4
level at which limit switch 4 responds
1 or 2
words
The formats of the cyclically communicated data content are specified in detail
in chapter 6.3. The measured values are offered optionally as a 16-bit integer,
32-bit integer or 32 bit float. The values are always scaled to physical size
with the number of decimal places of your choice. Information on whether the
16 bit format or a 32 bit format is used and on the number of decimal places is
defined in the parameter assignment message.
Typical combinations are predefined in the GSD file. If you require other
combinations you can expand the GSD file accordingly using the following
specifications.