PME-MP60/MP07
A0616-13.4 en
HBM
Operating status:
LED
colour
Status
Meaning
Measuring mode
Bus mode
Green
Steady
light
Ready to take measurements
CAN Operational (PDO transfer
possible)
Green
Flashing
Data is transferred via the
interface
-
Yellow
Steady
light
Ready to take measurements
CAN-bus Pre-operational (no PDO
transfer possible)
LED
colour
Status
Meaning
Remedy
Measuring mode
Bus mode
Red
Flashing
Measured value
overflow
LCD error
CAN transmission
error
Adapt measuring range
Restart
Red
Steady
light
Initialisation phase:
not ready to take
measurements yet,
calibration error
Original calibration
error
CAN-bus not ready for
communication (bus
OFF)
Please wait
Send PME to
manufacturer (HBM)