5.4 Example: measuring Md and N with torque
transducer T10F (24 V supply)
Both the T10F and the MP60 are set to symmetrical signals at the factory.
Torque measurement
Measurement preparation:
"
Use the standard cable md-cab149.
"
Connect the 24 V external supply voltage to terminal plug 4 as well (see
page 19).
Terminal plug 4
"
Connect the transducer (see Connection
A
"
No
DIP switch position changes!
Enter T10F characteristic values:
Group: TRANSDUCER
Unit:
kNm
Input range:
100 kHz
F2
OFF
Zero index (+)
OFF
Frq x 4
OFF
Direction of rotation
OFF
Switching threshold
2.5 V
Glitch filter
ON
Zero kHz
10
Zero kNm
0
NVal kHz
5 kH
z
NVal kNm
100 kNm
(see transducer identification plate)
Trans Err
OFF